MEASURING PARAMETER
7 Measuring parameter
7.1 Description
After switching on the
dynaROCK II
, the last employed combination of
measuring parameters is always active.
The measuring parameters in detail:
Impact direction : The impact direction is set-adjusted with the aid of the key
DIR in the measurement window.
Material :
The currently selected material.
Hardness scale : The current hardness scale into which the measured
values are converted as appropriate. In the measurement
window the hardness scale can be switched over with the
aid of the key SCALE, unless the test report printout or
the measured value memory are activated.
Rating :
Upper and lower limit for the rating GOOD are stored
here. If a measured value lies outside of these limits, an
acoustic signal is triggered with the measurement (2 short
tones). If the measured value lies within the limits, an
individual tone indicates the successful measurement.
If the value 0 is entered for upper and lower limit, no
verification of the measured value is implemented.
The value for the lower limit must of course be smaller
than that of the upper limit.
With active rating, the overshooting or undershooting of
the limits is displayed in the measurement window.
Statistics :
The number of measured values which should be
statistically evaluated without the use of the measured
value memory is defined here. If this number of measured
values is reached, the statistics window is opened
automatically (see 6.3.1).
24