CHAPTER 14
Specifications
129
Power
EXTERNAL DC POWER SUPPLY REQUIREMENTS
Used to charge the battery pack in the instrument
Voltage:
8–24 V DC, ripple voltage <20 mV
Current Requirement:
min. 1.5 A
Power Consumption:
<2.5 W, without battery
charging, <10 W when charging
Cable Connector:
LEMO Type FFA.00, positive at
centre pin
EXTERNAL AC MAIN SUPPLY ADAPTOR
Part No.
: ZG-0426
Supply Voltage:
100 to 120/200 to 240 VAC; 47 to
63 Hz
Connector:
2-pin IEC 320
BATTERY PACK
Rechargeable Li-Ion battery
Part No.:
QB-0061
Voltage:
3.7 V
Capacity:
5200 mAh nominal
Typical Operating Time:
>10 h (screen backlight
dimmed); >7 h (full screen backlight)
Battery Cycle Life:
>500 complete charge/discharge
cycles
Battery Indicator:
Remaining battery capacity and
expected working time may be read out in % and time
Battery Fuel Gauge:
The battery is equipped with a
built-in fuel gauge, which continuously measures and
stores the actual battery capacity in the battery unit
Charge Time:
In instrument, typically 10 hours from
empty at ambient temperatures below 30°C. To protect
the battery, charging will be terminated completely at
ambient temperatures above 40°C. At 30 to 40°C
charging time will be prolonged. With External Charger
ZG-0444 (optional accessory), typically five hours
Note:
It is not recommended to charge the battery at
temperatures below 0°C (32°F) or over 50°C (122°F).
Doing this will reduce battery lifetime
CLOCK
Back-up battery powered clock. Drift <0.45 s per 24-
hour period
Storage
INTERNAL FLASH-RAM (NON-VOLATILE)
20 Mbyte for user setups and measurement data
EXTERNAL MEMORY CARD
Secure Digital (SD and SDHC) Card:
For store/recall
of measurement data
Compact Flash (CF) Card:
Hardware versions 1 – 3
only. For store/recall of measurement data
USB MEMORY STICK (Hardware Version 4 only)
For store/recall of measurement data
Environmental
WARM-UP TIME
From Power Off:
<2 minutes
From Standby:
<10 seconds
TEMPERATURE
IEC 60068–2–1 & IEC 60068–2–2:
Environmental
Testing. Cold and Dry Heat
Operating Temperature:
–10 to +50°C (14 to 122°F),
<0.1 dB
Storage Temperature:
–25 to +70°C (–13 to +158°F)
HUMIDITY
IEC 60068–2–78:
Damp Heat: 90% RH (non-
condensing at 40°C (104°F))
Effect of Humidity:
< 0.1 dB for 0% < RH < 90% (at
40°C (104°F) and 1 kHz)
MECHANICAL
Environmental Protection:
IP 44
Non-operating:
• IEC 60068–2–6: Vibration: 0.3 mm, 20 m/s
2
,
10 – 500 Hz
• IEC 60068–2–27: Shock: 1000 m/s
2
• IEC 60068–2–29: Bump: 4000 bumps at 400 m/s
2
WEIGHT AND DIMENSIONS
650 g (23 oz.) including rechargeable battery
300
93
50 mm (11.8
3.7
1.9
) including
preamplifier and microphone
User Interface
LANGUAGE
User Interface in Catalan, Chinese (People’s Republic
of China), Chinese (Taiwan), Croatian, Czech, Danish,
English, Flemish, French, German, Hungarian,
Japanese, Italian, Korean, Polish, Portuguese,
Romanian, Russian, Serbian, Slovenian, Spanish,
Swedish and Turkish
HELP
Concise context-sensitive help in English, French,
German, Italian, Japanese, Korean, Polish,
Portuguese, Romanian, Serbian, Slovenian and
Spanish
UPDATE OF SOFTWARE
Update to any version using BZ-5503 through USB or
update via Internet:
• Hardware versions 1 – 3: the latest version only
• Hardware version 4: any version from 4.0 and up
WEB PAGE
Connect to instrument via Internet browser supporting
JavaScript. The connection is password protected
Содержание 2250 Light
Страница 2: ......
Страница 10: ......
Страница 48: ...38 2250 Light User Manual...
Страница 58: ...48 2250 Light User Manual...
Страница 116: ...106 2250 Light User Manual...
Страница 126: ...116 2250 Light User Manual...
Страница 136: ...126 2250 Light User Manual...
Страница 168: ...2250 Light User Manual 158...
Страница 205: ......
Страница 206: ......
Страница 207: ......