CHAPTER 11
Logging (Optional Module)
99
Fig.11.1
The Setup screen
5) Under the
Measurement Control
parameters, set the
Preset Logging Time
and
Logging
Period
as required. Set
Synchronize
with clock
to
Yes
if you want the logging to
synchronise with whole minutes or hours. For example, if
Logging Period
is set to
1 minute and you start the measurement at 8:12:33, then the first logging interval will be
from 8:12:33 to 8:12:59 (27 seconds), the second will be from 8:13:00 to 8:13:59
(60 seconds), etc. Set
Synchronize with clock
to
No
if you want every logging interval to
be the specified Logging Period exactly. Set
Charge Injection Calibration
to
On
, if you
want to validate the complete measurement chain by performing CIC at the start and end of
the logging measurement, refer to section 5.6.
6) Under the
Logged Broadband
parameters, choose which broadband parameters you want
to log in accordance with the
Measurement Control
parameters. You can choose to log
Full
Statistics
per
Logging Period
or not. You can also choose to log
all
the measured
Broadband
Parameters
or a
Selected
number of parameters. If you choose
Selected
then you can
specify up to 10 parameters.
7) The
Logged Broadband (100 ms)
parameter allows you to log L
Aeq
(with an elapsed time
of 100 ms and a logging period of 100 ms), L
AF
and/or L
AS
(with a logging period of
100 ms) irrespective of the other logging parameters.
8) The
Logged Spectrum
a
parameter allows you to choose which spectra to log. You can
specify to log
All
,
None
or up to 3
Selected
spectra.
9) Under
Markers
you can specify the names of the five available markers. The markers are
predefined as follows:
•
Marker 1: ‘Exclude’ – controlled by pressing the
Back-erase
pushbutton
•
Marker 2: ‘Event 1’ – controlled by pressing the
Manual
Event
1
pushbutton
•
Marker 3: ‘Event 2’ – controlled by pressing the second
Manual Event 2
pushbutton
•
Marker 4: ‘Marker 4’
•
Marker 5: ‘Marker 5’
a.
Requires 1/1- or 1/3-octave Frequency Analysis Software for 2250 Light BZ-7131/32
Содержание 2250 Light
Страница 2: ......
Страница 10: ......
Страница 48: ...38 2250 Light User Manual...
Страница 58: ...48 2250 Light User Manual...
Страница 116: ...106 2250 Light User Manual...
Страница 126: ...116 2250 Light User Manual...
Страница 136: ...126 2250 Light User Manual...
Страница 168: ...2250 Light User Manual 158...
Страница 205: ......
Страница 206: ......
Страница 207: ......