Documentation for SONOREX CNp 28-2
BANDELIN electronic GmbH & Co. KG, Heinrichstraße 3-4, 12207 Berlin, Germany Page 32 from 55
6.6
Carry out rinsing programme
After completing the cleaning programme, place the basket with the items to be cleaned in the
rinsing bath, close the lid and start the rinsing programme.
6.7
Complete the cleaning process
After finishing the rinsing programme, remove the parts and dry them. Close the lids of the baths
to reduce evaporation.
6.8
Configure cleaning programme
Select the desired cleaning procedure by touching the ultrasonic, heating and vacuum selection.
The user login for entering the default password: 12345 appears. After entering the password,
configure the programme duration by touching the time display. Configure the set temperature by
touching the temperature scale.
6.9 Configure parameters
Select the "Parameter" icon in the settings menu. The time specification corresponds to seconds
multiplied by 1000. The times of the vacuum cycle can be set. The start vacuum duration (default
value 20,000) is defined by the initial vacuum time after programme start. The vacuum pulse
length (default value 8000) is defined by the duration of the activated vacuum during the process.
The ventilation length (default value 400) defines the duration of ventilation of the cleaning
chamber during the process.
Figure 1: Configure parameters