Original - Operating Instructions SONOREX CNp 28-2
BANDELIN electronic GmbH & Co KG, Heinrichstraße 3-4, 12207 Berlin, Germany Page 27 from 55
5.2.3
Connect unit
Connect waste water and suction air outlet
-
Check whether the on-site waste water pipe has sufficient ventilation.
If a lifting system is provided on site, ventilation of the lifting system chamber must be
ensured. The waste water and suction air outlets must always be open.
-
Fit the pipe bracket to the waste water and suction air outlets.
-
Connect the waste water and suction air outlet to the on-site waste water pipe using the
connecting hose supplied.
There must be no blockages in the waste water pipe. Ensure a constant slope of the pipe from the
unit to the waste water system.
Connect power and compressed air supply
-
Observe the wiring diagrams and compliance of the existing mains voltage with the mains
voltage specified for operating the unit.
-
Observe the connection data in the technical data and in the wiring diagram.
-
Lay the connection cable with strain relief.
-
Insert the plug of the unit into the corresponding socket. Please refer to the technical data for
the plug size.
-
Connect the compressed air line to the house network.
CAUTION