X.01
X100
X.1 X1000
X1 X10,000
X10 X100,000
The following screen will appear:
CHART REVOLUTION SPEED
1 REVOLUTION IN 1 DAY
This screen allows the selection of the chart rotation
speed. The following speeds are available:
1 REVOLUTION IN 2 MINUTES
1 REVOLUTION IN 24 MINUTES
1 REVOLUTION IN 1 HOUR
1 REVOLUTION IN 1 DAY
1 REVOLUTION IN 7 DAYS
1 REVOLUTION IN 31 DAYS
- - - - -
- - -
The two dash line positions are for future chart
revolution speeds. Press the UP or DOWN arrow keys to
select the desired chart speed and press the ENTER key.
The following screen will appear:
RECORDER DAMPING
DAMPING TIME= 6 SECS.
This screen allows the selection of the amount of
damping in the response of the pen movement. The
damping times available are:
NONE
6
16 40
100
240
600
These times are in seconds. Press the UP or DOWN
arrow keys to select the desired value and then the
ENTER key. The following screen will appear:
- FUNCTION SELECTION -
POWER=> LINEAR
This screen allows the selection of linearization
functions for use with differential pressure transmitters,
open channel flow measurement devices or any definable
function. Check the output specification of the device that
is being connected to the recorder to determine the proper
power selection. The functions available are:
POWER=>LINEAR
POWER=>1.55
POWER=>1.58
POWER=>1.95
POWER=>1/2
POWER=>3/2
POWER=>5/2
POWER=>OPTION
The linear function is used where the transmitter output
is proportional to flow.
The 1.55 power is used with 1", 2", 3", 24" Parshall
flumes and all Lagco flumes.
The 1.58 power is used with a 6" Parshall flume.
The 1.95 power is used with the Badger Meter Manhole
flumes.
The 1/2 power is used with differential pressure
transmitters where the output does not have the square
root extracted.
The 3/2 power is used with Suppressed Rectangular
weirs and Cipoletti weirs.
The 5/2 power is used with all V-Notch weirs.
The OPTION power allows a 16 point HQ curve to be
programmed into the recorder. Consult the factory for
programming details.
Press the UP or DOWN arrow keys to select the desired
function and then the ENTER key. The following screen
will appear:
INPUT SELECTION
CHANNEL=ANALOG #1
This screen allows the selection of the desired input
channels to be used as the input to the recorder. There are
four channels available to connect the input(s) signal(s):
ANALOG #1, ANALOG #2, ANALOG #3, AND
FREQUENCY.
The ANALOG channels can accept 0-20 mA, 4-20 mA,
or 1-5 VDC. The recorder must be set up at the factory
for either a current or a voltage input. These channels can
be selected individually or summed in any combination.
The frequency channel is used for pulsed inputs. The
operating range is 0-1000 pps to 0-150 ppm. Press the UP
or DOWN arrow keys to select the desired input channel
and then the ENTER key.
10