Manual Part Number:
5-06-4900-65-0
Revision Level: A
ECO 8541
File Name: MN-A-0177
Page 17
6. Verify Outputs:
If the Model 5124BE-N1-WO analyzer indicates between 20.8% and 21.0%
after warm-up, and all the first-time power-up verifications have been completed, Check the
fault, temperature OK, and oxygen alarm relay contacts to ensure that they are wired
properly and functioning normally. Also, check the analog current, analog voltage, and
range ID voltage outputs to ensure that they are functioning normally.
7. Apply the Gas to Be Analyzed:
Apply the gas that is to be analyzed to the analyzer. Verify
that the oxygen concentration readout on the RS-232 terminal, relay contact outputs, the
analog outputs and the Range ID outputs are wired properly and functioning normally.
B. Normal Power Up and Power Down Instructions
After the instructions for the first-time power up has been completed, the Model 5124BE-N1-WO
analyzer can be run continuously. For periods when the analyzer is not in use, it may be powered
down. Wait at least 15 minutes after power down to ensure that the sensor is cool before
reapplying power to the analyzer.
C. Oxygen Alarm Level Settings
If the user's process requires adjustments to either oxygen alarm level, follow adjustment
procedure in Appendix C.