Manual Part Number:
5-06-4900-65-0
Revision Level: A
ECO 8541
File Name: MN-A-0177
Page 4
Step 1
– Analyzer Installation
WARNING:
The internal components of the Model 5124BE-N1-WO analyzer and
the remote sensor are fragile. Do not drop the analyzer or the remote sensor during
installation. Do not expose the unit to shock or vibration during installation. Failure to
comply may damage the fragile components inside the analyzer and remote sensor
may result in loss of warranty.
The Model 5124BE-N1-WO analyzer is designed to be mounted flush to the surface of an
equipment control panel. Observe the following guidelines when installing the analyzer:
•
Allow ample space at the rear of the analyzer for easy access to all electrical and
pneumatic connections.
•
Be sure that the analyzer will not be exposed to water, adverse temperature, or shock.
•
Be sure that the analyzer has free air flow around the vent slots on the chassis to prevent
overheating.
PANEL CUTOUT & DRILL PATTERN
4 PLACES
6.62" 16.81cm
6.20" 15.50cm
2.75"
6.99cm
2.91"
7.39cm
.189 DIA. HOLE (.429cm)
Figure 3
– Analyzer Panel Cutout Diagram