Manual Part Number:
5-06-4900-65-0
Revision Level: A
ECO 8541
File Name: MN-A-0177
Page 7
Step 3
– Electrical Installation
CAUTION:
Be sure all mains power supply is off before attempting the electrical
connections to the analyzer and remote sensor. DO NOT WORK WITH LIVE WIRES!
Do not leave any exposed wire at the connectors or cables. Before applying power,
ensure terminal blocks are fully inserted into the mating connector at the analyzer.
DANGER:
The Model 5124BE-N1-WO analyzer and its external sensor are not rated
intrinsically safe or explosion proof. Be certain that no flammable gases are present in the areas
where the analyzer and the external sensor will be installed.
CAUTION:
The Model 5124BE-N1-WO analyzer and external housings are not rated
waterproof. Do not mount the analyzer or the sensor in an area where it may contact water or
other liquid elements
.
WARNING:
The Model 5124BE-N1-WO is not equipped with a circuit breaker. A
disconnecting switch or a circuit breaker must be installed in series with the mains power to the
analyzer.
All electrical connections to the Model 5124BE-N1-WO analyzer is to be made to the three terminal
blocks and sensor interface socket located at the back of the analyzer chassis. A label depicting
the terminal block arrangement and the location of the bayonet connector is affixed to the top of
the chassis as reference. The terminal blocks may be removed for wiring.
The electrical connection to the Model 5124BE-N1-WO external sensor is to be made to the
analyzer interface socket. Use the sensor interface cable provided by Neutronics.