![Bacharach MGS-400 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/bacharach/mgs-400/mgs-400_user-manual_3332627049.webp)
1100-2294 Rev 1
48
1138
Gas Type Text Char 5,6
Gas Type Characters 5 & 6 (10 character gas string =
"XXXXXXXXXX")
16-bit Unsigned
Integer
1139
Gas Type Text Char 7,8
Gas Type Characters 7 & 8 (10 character gas string =
"XXXXXXXXXX")
1140
Gas Type Text Char 9,10
Gas Type Characters 9 & 10 (10 character gas string =
"XXXXXXXXXX")
1141
Sensor Module SID Char 1,2
SID Characters 1 & 2 (8 character UID string =
"XXXXXXXX")
1142
Sensor Module SID Char 3,4
SID Characters 3 & 4 (8 character UID string =
"XXXXXXXX")
1143
Sensor Module SID Char 5,6
SID Characters 5 & 6 (8 character UID string =
"XXXXXXXX")
1144
Sensor Module SID Char 7,8
SID Characters 7 & 8 (8 character UID string =
"XXXXXXXX")
1145
Sensor Controller UID Char
1,2
UID Characters 1 & 2 (8 character UID string =
"XXXXXXXX")
1146
Sensor Controller UID Char
3,4
UID Characters 3 & 4 (8 character UID string =
"XXXXXXXX")
1147
Sensor Controller UID Char
5,6
UID Characters 5 & 6 (8 character UID string =
"XXXXXXXX")
1148
Sensor Controller UID Char
7,8
UID Characters 7 & 8 (8 character UID string =
"XXXXXXXX")
1149
Alias Text Char 1,2
Alias Characters (16 character alias string =
"XXXXXXXXXXXXXXXX")
1150
Alias Text Char 3,4
1151
Alias Text Char 5,6
1152
Alias Text Char 7,8
1153
Alias Text Char 9,10
1154
Alias Text Char 11,12
1155
Alias Text Char 13,14
1156
Alias Text Char 15,16
1157
Software Version Sensor
Major
Major software version level (XX in firmware XX.YY.ZZ
format)
1158
Software Version Sensor
Minor
Minor software version level (YY in firmware XX.YY.ZZ
format)
1159
Software Version Sensor
Bug fix
Bug fix software version level (ZZ in firmware
XX.YY.ZZ format)
1160
Temperature Lower Limit
(°C)
Set Temperature Fault flag when Temp < Temperature
Lower Limit
16-bit Signed
Integer
1161
Temperature Upper Limit
(°C)
Set Temperature Fault flag when Temp > Temperature
Upper Limit
6.4.3 Integration - General System Setup
Holding Register
(Function 03/06 Read / Write)
🔒
2100
Parameter Unlock
Writing the correct unlock code allows an external
controller to change system parameters (0000-9999)
16-bit Unsigned
Integer
2101
🔒
RS-485 Node Address
Modbus address 1-247 (if hardware override - write
exception / read Modbus switch state)
2102
🔒
Baud Rate
0 = 9600 Baud; 1 = 19200 Baud (if hardware override -
write exception / read dip8)
Boolean
Содержание MGS-400
Страница 1: ...1100 2294 Rev 1 1...