4.2.3 Instrument Configuration
For security, access to configuration and calibration options are restricted to authorized users
only. Access to these functions require use of an unlock code. To unlock instrument
configuration:
Configure Tab
When prompted, enter unlock code to access device configuration.
(
The instrument’s default code is “1234”)
. Instrument will remain unlocked until
Bluetooth
®
connection has ended.
WARNING:
Default alias, passkey and unlock code can be changed via the
MGS-
400 App’s configuration menu. Default values should be changed after
instrument installation for security purposes.
4.2.3.1 Change Alias
To allow easy identification of a given instrument, an alias can be assigned to each instrument.
This alias is displayed when searching for an instrument via Bluetooth
®
, on calibration cert and
in home tab. To set alias:
Configure Tab
Alias, Enter required alias for instrument, select OK.
Instrument must be restarted for change to take effect. Home Tab
Restart, will reboot
device.
Reconnect to instrument to confirm alias has been updated.
4.2.3.2 Change Unlock Code
To prevent unauthorized access to instrument configuration and calibration, default instrument
unlock code should be changed during commissioning. To change unlock code:
Configure Tab
Modbus Unlock Code, enter new 4-digit unlock code for instrument,
select OK.
Instrument must be restarted for change to take effect. Home Tab
Restart, will reboot
device.
Reconnect to instrument to confirm unlock code has been updated.
IMPORTANT:
If custom unlock code is forgotten, unlock code may be reset to
default value (1234) by resetting system to factory defaults. Refer to section
4.1.6 for system reset procedure. Note system reset will return all custom system
configurations to defaults
4.2.3.3 Change Bluetooth Passcode
To prevent unauthorized access to instrument status, default instrument Bluetooth
®
passcode
code should be changed during commissioning. To change Bluetooth
®
passcode:
Configure Tab
Bluetooth Passcode, enter new 6-digit passcode for instrument, select
OK.