1100-2294 Rev 1
33
Analog Span Range
Analog span range scales the FSD
(full-scale deflection)
of the analog output. The selected
range determines the equivalent gas measurement at the analog output maximum range.
Example: R134A 1000 ppm, 0-5V analog output. If Analog Span Range is set to 20%, the full
analog output range only covers the first 20% of the gas measurement range, i.e 0-200 ppm will
output 0-5V, above 200 ppm the output will be truncated to 5V.
Note, sensor resolution stays at the value for the max range.
Adjustment is limited to between 20%-100% FSD, Default is set to 100%. To set analog span
range:
Configure Tab
Outputs
Analog Span Range, use slider to set desired range
Alternatively, tap “Analog Span Range (X.X%)” text and enter specific range required.