13
Safety instructions
4.1.9
Enteral nutrition
The Infusomat compact
plus
P
can be used
for enteral nutrition.
Q
Do not use enteral fluids for the intra-
venous infusion. This would lead to a
risk of severe injury or death for the
patient.
Q
Only use ENFit feed systems that have
been designed and designated for
enteral nutrition.
4.2
Electrical connection
Q
Do not use the device if the plug has
visible damage.
Q
Do not use an extension cable that
has not been approved for use with
device. Position the power cable so
that it does not present a trip hazard.
4.3
Safety standards
Q
The Infusomat compact
plus
P complies
with all safety standards for medical
electrical devices in accordance with
–
IEC 60601-1:2005
–
IEC 60601-1:2005/AMD1:2012
–
IEC 60601-1-6:2010
–
IEC 60601-1-6:2010/AMD1:2013
–
IEC 60601-1-8:2006
–
IEC 60601-1-8:2006/AMD1:2012
–
IEC 60601-1-12:2014
–
IEC 60601-2-24:2012
Q
The EMC limits (electromagnetic
compatibility) according to
–
IEC 60601-1-2:2007
–
IEC 60601-2-24:2012
are complied with.
Hinweis DTP 05-20
Ende Kapitel 4.1.8:
Achtung!
Die letzten Worte aus
der Anweisung der PDF
(in Klammern, BBM...)
wurden von mir nicht
eingesetzt. Wie ich denke
ist dies nur ein interner
Hinweis.
--> Bitte bestätigen.
Vollständiger Text der
Anweisung aus dem PDF:
Where several infusion
lines are connected
on one single vascular
access, the possibility
of the lines exerting a
mutual influence over
each other cannot be
excluded
(BBM Applica-
tion notes on parallel
infusion 38910004).
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 13
15.10.2020 17:47:25