55
Instructions for use for accessories
The device has two different staff call operating modes:
Switched off
Static
without
off alarm *)
Dyn.
without
off alarm *)
Alarm
Opera-
tion
Alarm
Opera-
tion
Switched off
Switched on
Operating alarm
1 sec.
* In “static without off alarm” mode, the staff call can be disabled by pressing the OK key.
Technical data
Mode:
Status
red LED
Status Changeover Contact
(wire color white - green)
Status Changeover Contact
(wire color white - brown]
Operating:
off
closed
opened
Alarm:
on
opened
closed
Accessory
Parameter
Value
Interface lead 12 V CP
(8718020)
Dimensions in mm
(W x H x D)
Cable length in mm
Approx. 75 mm x 43 mm x 15 mm
Approx. 2568 mm
Interface lead staff call CP
(8718030)
Dimensions in mm
(W x H x D)
Cable length in mm
Approx. 75 mm x 43 mm x 15 mm
Approx. 2553 mm
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 55