32
Operation
The main menu is displayed.
Q
Select
Dose calculation
and press
OK
to confirm.
Q
Select active substance unit
and
press
OK
to confirm.
Q
Enter active substance quantity
and
press
OK
to confirm.
Q
Enter volume
and press
OK
to con-
firm.
The “Calculate Using:” screen is
displayed.
Calculating without patient data
The delivery rate is calculated without any
patient data being entered.
Q
Select
No patient data
and press
OK
to confirm.
Q
Select
dose unit
and press
OK
to
confirm.
Q
Enter
dose
.
Note:
Pressing the OK key brings up the
Overview screen.
Q
Check the plausibility of the displayed
values.
Q
Start the infusion with the
Start/Stop
key
.
Soft limits
Soft limits for rate/dose/bolus volume
and bolus rate can also be stored in the
database. These can be exceeded with
no restriction but the following message
appears on the display.
The following symbols that describe the
status of the pump with regard to the soft
limits are described:
Symbol
Meaning
No symbol
Infusion is within the
soft limits
Infusion is above the
upper soft limit
Infusion is below the
lower soft limit
8.10
Calculating the dose
The
Dose calculation
function is used to
calculate the delivery rate in ml/h based on
the dose parameters entered.
Q
Infusion line inserted and selected
Q
Press the
Menu key.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 32
15.10.2020 17:47:29