28
Operation
8.5
Setting the infusion values
Q
Infusion line inserted and selected
Note:
Depending on the last therapy, the
pump starts either when the delivery rate is
entered or when a drug is selected.
8.5.1
Entering the delivery rate
Q
Enter the
delivery rate
using the arrow
keys.
Q
Start the infusion with
the Start/Stop
key
.
- or -
Q
Press
OK
to confirm the rate.
The
Overview
screen is displayed.
Q
Select Vol.
/
Time
and press
OK
to con-
firm.
Q
Enter the volume
or
time limit
and
press
OK
to confirm.
Any values still missing are automati-
cally calculated and displayed.
Note:
In addition to the volume and time
limit, the infusion rate can also be adjusted
in the
Overview
screen.
Q
Start the infusion with the
Start/Stop
key
.
8.4
Use with Intrafix® SafeSet
Note:
If Intrafix® SafeSet is used, a VTBI is
not required.
A special airtight filter membrane acts as
a barrier and helps prevent air entering the
line. If the fluid level reaches the membrane,
the upstream alarm is triggered and the
pump stops so that no air passes through
the AirStop filter. For this reason, no addi-
tional priming is required when quickly
changing to the next container.
In the case of an upstream alarm, the
upstream sensor detects a low pressure
in the infusion line between the pump and
the drip chamber. Therefore, always check
whether the roller clamp is open, the line
is bent or the bag and/or drip chamber is
empty.
The pump should not be restarted until the
upstream alarm has been resolved.
If the pump is started too many times
without the problem having been rectified,
the upstream sensor calibrates to the low
pressure in the line at the time and air can
pass through the air stop membrane. In this
case, ensure that the drip chamber is
refilled, then open the flap to recalibrate
the upstream sensor.
38932401_Infusomat_compactPlusP_EN_IFU_1020_v5.indd 28
15.10.2020 17:47:28