27
Operation
Q
The pump will calibrate the infusion
line.
Entries can be made during the
calibration. However, the infusion can
only be started after the infusion line
has been successful calibrated.
Line calibration
ml/h
Note:
Only when the pump is switched on,
the door is completely closed, and the dis-
posable item is correctly inserted, will the
pump have control over the disposable item
and thus protect against free flow.
Note:
Do not stretch or twist the infusion
line during insertion.
Note:
Use of a clamp above the drip chamber
is dependent on the application (e. g. CytoSet)
and is the responsibility of the user.
The status of the clamp (open/closed) is not
monitored by the pump.
8.3
Priming the infusion line
Note:
The “Prime infusion line” function is
started after an air alarm. It enables the line
to be primed by the pump without having
to remove the line. The line must be discon-
nected from the patient for this purpose.
CAUTION! Ensure that the line guide
is straight and avoid sagging/bending,
otherwise the line may be damaged
and could cause incorrect dosing.
Q
Close the pump door while simultane-
ously operating the door lever.
Do not damage the infusion line.
Q
Follow the instructions on the screen
and open the roller clamp.
The “Select infusion line” message will
be displayed.
Q
Select “Line type” and press OK to
confirm. Make sure that the line type
displayed is the same as the inserted
line.