4
Cervical – Navigated Instruments
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested and approved effec-
tiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced
and checked at regular intervals.
Manual pre-cleaning with a brush
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
►
Note the information on appropriate cleaning brushes and disposable
syringes, see Validated cleaning and disinfection procedure.
Phase I
►
Fully immerse the product in the cleaning/disinfectant for at least 15
min. Ensure that all accessible surfaces are moistened.
►
Clean the product with a suitable cleaning brush in the solution until
all discernible residues have been removed from the surface.
►
If applicable, brush through non-visible surfaces with an appropriate
cleaning brush for at least 1 min.
►
Mobilize non-rigid components, such as set screws, links, etc. during
cleaning.
►
Thoroughly rinse through these components with the cleaning disin-
fectant solution (at least five times), using a disposable syringe.
Phase II
►
Rinse/flush the product thoroughly (all accessible surfaces) under run-
ning water.
►
Mobilize non-rigid components, such as set screws, joints, etc. during
rinsing.
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-