16
Aesculap
®
S
4
Cervical – Navigated Instruments
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with
immersion disinfection
■
FW652R–FW658R
■
FW660R
■
FW663R–FW665R
■
FW688R–FW689R
■
Cleaning brush: 30 mm/
∅
: 4.5 mm, e.g. TA011944 and cleaning
brush: 20 mm/
∅
: 2.5 mm, e.g. TE654202 and cleaning brush:
50 mm/
∅
: 10 mm, e.g. TA007747
■
20 ml disposable syringe
■
Keep working ends open for cleaning.
■
When cleaning products with movable hinges, ensure that these are
in an open position and, if applicable, move the joint while cleaning.
■
Drying phase: Use a lint-free cloth or medical compressed air
Chapter Manual cleaning/disinfection and sub-
chapter:
■
Chapter Manual cleaning with immersion
disinfection
Manual cleaning with
ultrasound and immer-
sion disinfection
■
FJ985R
■
FW067R
■
FW165R
■
FW661R
■
Cleaning brush: 30 mm/
∅
: 4.5 mm, e.g. TA011944 and cleaning
brush: 20 mm/
∅
: 2.5 mm, e.g. TE654202 and cleaning brush:
50 mm/
∅
: 10 mm, e.g. TA011327
■
20 ml disposable syringe
■
Keep working ends open for cleaning.
■
When cleaning products with movable hinges, ensure that these are
in an open position and, if applicable, move the joint while cleaning.
■
Drying phase: Use a lint-free cloth or medical compressed air
Chapter Manual cleaning/disinfection and sub-
chapter:
■
Chapter Manual cleaning with ultrasound
and immersion disinfection
Mechanical alkaline
cleaning and thermal
disinfection
■
FW657R
■
FW663R
■
FW688R–FW689R
■
Connect components with lumens and channels directly to the rins-
ing port of the injector carriage.
■
Keep working ends open for cleaning.
■
Place products in the tray with their hinges open.
Chapter Mechanical cleaning/disinfecting and
sub-chapter:
■
Chapter Mechanical alkaline cleaning and
thermal disinfecting
Manual pre-cleaning
with brush and subse-
quent mechanical alka-
line cleaning and ther-
mal disinfection
■
FW067R
■
FW165R
■
FW652R
■
FW655R–FW656R
■
FW658R
■
FW660R
■
FW664R–FW665R
■
Cleaning brush: 30 mm/
∅
: 4.5 mm, e.g. TA011944 and cleaning
brush: 20 mm/
∅
: 2.5 mm, e.g. TE654202 and cleaning brush:
50 mm/
∅
: 10 mm, e.g. TA007747
■
20 ml disposable syringe
■
Place the product in a tray that is suitable for cleaning (avoiding rins-
ing blind spots).
■
Keep working ends open for cleaning.
■
Place products in the tray with their hinges open.
Chapter Mechanical cleaning/disinfection with
manual pre-cleaning and sub-chapter:
■
Chapter Manual pre-cleaning with a brush
■
Chapter Mechanical alkaline cleaning and
thermal disinfecting