Validated cleaning and disinfection procedure
For mechanical cleaning:
Note
Use appropriate (neutral, enzymatic and mild-alkaline) detergents on this product with its aluminum components.
Manual cleaning/disinfection
►
Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.
►
After manual cleaning/disinfection, check visible surfaces visually for residues.
►
Repeat the cleaning /disinfection process if necessary.
Manual cleaning with ultrasound and immersion disinfection
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
►
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
►
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible
surfaces are immersed and acoustic shadows are avoided.
►
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
►
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
►
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
►
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
►
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
►
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
►
Drain any remaining water fully.
Phase III
►
Fully immerse the product in the disinfectant solution.
►
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
►
Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.
Phase IV
►
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
►
Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.
►
Rinse lumens with an appropriate disposable syringe at least five times.
►
Drain any remaining water fully.
Phase V
►
Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with ultrasound and brush
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
►
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
►
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible
surfaces are immersed and acoustic shadows are avoided.
►
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
►
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
►
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
►
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
►
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
►
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Mechanical neutral or mild alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
*Recommended: BBraun Helimatic Cleaner neutral
►
Check visible surfaces for residues after mechanical cleaning/disinfecting.
►
Repeat the cleaning/disinfecting process if necessary.
Inspection, maintenance and checks
►
Allow the product to cool down to room temperature.
►
After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational,
and free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured compo-
nents).
►
Dry the product if it is wet or damp.
►
Repeat cleaning and disinfection of products that still show impurities or contamination.
►
Check that the product functions correctly.
►
Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service.
►
Assemble dismountable products, see Assembling.
►
Check for compatibility with associated products.
Packaging
►
Appropriately protect products with fine working tips.
►
Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected.
►
Pack trays appropriately for the intended sterilization process (e.g. in sterile Aesculap containers).
►
Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Validated procedure
Specific requirements
Reference
Manual cleaning with
ultrasound and immer-
sion disinfection
■
Cleaning brush: 30 mm/
∅
: 4.5 mm,
e.g. TA no. 011944
■
20 ml disposable syringe
■
Mount jaws protection on the prod-
uct.
■
Keep working ends open for cleaning.
■
When cleaning products with mov-
able hinges, ensure that these are in
an open position and, if applicable,
move the joint while cleaning.
■
Drying phase: Use a lint-free cloth or
medical compressed air
Chapter Manual cleaning/disinfection
and sub-chapter:
■
Chapter Manual cleaning with
ultrasound and immersion disin-
fection
Manual pre-cleaning
with ultrasound and
brush and subsequent
mechanical neutral or
mild alkaline cleaning
and thermal disinfecting
■
Cleaning brush: 30 mm/
∅
: 4.5 mm,
e.g. TA no. 011944
■
20 ml disposable syringe
■
Mount jaws protection on the prod-
uct.
■
Place the instrument in a tray that is
suitable for cleaning (avoiding rinsing
blind spots).
■
Keep working ends open for cleaning.
■
Place instruments in the tray with
their hinges open.
Chapter Mechanical cleaning/disin-
fection with manual pre-cleaning and
sub-chapter:
■
Chapter Manual pre-cleaning with
ultrasound and brush
■
Chapter Mechanical neutral or
mild alkaline cleaning and thermal
disinfecting
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I
Ultrasonic
cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
II
Intermediate
rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I
Ultrasonic
cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concen-
trate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
Phase
Step
T
[°C/°F]
t
[min]
Water
quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
Neutral:
■
Concentrate:
–
pH neutral
–
<5 % anionic surfactant
■
0.5 %* working solution
Mildly alkaline:
■
Concentrate:
–
pH = 9.5
–
<5 % anionic surfactant
■
0.5 % solution
III
Intermediate
rinse
>10/50
1
FD-W
-
IV
Thermal disin-
fecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning
and disinfection device