background image

AB-7050 

Specifications 

Item Specification 

Temperature 0 

°

C to 60 

°

Measuring range 

Humidity 

0 %RH to 100 %RH (at 15 

°

C 35 

°

C) 

RTD 
(Pt100) 

±

 0.3 

°

C (at 0 

°

C to 60 

°

C)  

* 1mA applied current, 0.15 m/s downflow 
for Models TY7043Z0P0X(-1), TY7053Z0P00(-1), and HTY7043 series 

Temperature

RTD 
(Pt1000) 

±

 0.5 

°

C (at 0 

°

C to 60 

°

C)  

* 0.1-0.3mA applied current, 0.15 m/s downflow 
for Model TY7043Z0K00(-1) 

Measuring accuracy 

Humidity 

±

 3 %RH (within 30 %RH to 70 %RH range, at 25 

°

C), 

±

 5 %RH (within 20 %RH to 80 %RH

range, at 15 

°

C to 35 

°

C)  

for Models HY7043 series and HTY7043 series 
100 

 / 0 

°

C (RTD (Pt100) conforming to JIS C 1604 Class A)  

for Models TY7043Z0P0X(-1), TY7053 series, HTY7043 series 

Temperature 

1000 

/ 0 

°

C (RTD (Pt1000)equivalent to JIS C 1604 Class A)  

for Model TY7043Z0K00(-1) 
1 V DC to 5 V DC (linear to 0 %RH to 100 %RH)  
* Min. 50 k

 input impedance of the controller in connection 

for Models HY7043T1000(-1), HTY7043T1P00(-1) 

Output signal 

Humidity 

0 V DC to 10 V DC (linear to 0 %RH to 100 %RH)  
* Min. 50 k

 input impedance of the controller in connection 

for Models HY7043T6000(-1), HTY7043T6P00(-1) 
Max. 4.5 min. (at 0.15 m/s air velocity)  
for Models TY7043 series and HTY7043 series (Neosensor) 

Temperature 

Max. 2.5 min. (reference value) (at 0.15 m/s air velocity)  
for Model TY7053 series (Neosensor high sensitive type) 

Time constant  
(at 0.15 m/s wind 
velocity) 

Humidity 

Max. 40 sec. (at 0.15 m/s air velocity)  
for Models HY7043 series and HTY7043 series (Neosensor) 

Power supply 

24 V AC +10/-15 % (50 Hz/60 Hz), 24 V DC 

±

 10 % 

for Models HY7043 series, HTY7043 series 

Power consumption 

0.15 VA (24 V AC power), 100 mW (24 V DC power) 
for Models HY7043TX000(-1), HTY7043TXP00(-1) 

Insulation resistance 

Min. 500 V DC, 20 M

 

Withstand voltage 

500 V AC applied for 1 min. with max. of 1 mA leakage current  
(between housing and terminals) 

Environmental conditions 

Rated operating condition

Limit operating condition Transport/storage condition

Ambient temperature 

°

C to 60 

°

C -10 

°

C to 60 

°

C -20 

°

C to 70 

°

For temperature 
sensing 

Ambient humidity 

0 %RH to 100 %RH 
(non-condensing) 

0 %RH to 100 %RH 
(non-condensing) 

5 %RH to 95 %RH 
(non-condensing) 

Ambient temperature 

15 

°

C to 35 

°

C -10 

°

C to 60 

°

C -20 

°

C to 70 

°

For humidity 
sensing 

Ambient humidity 

20 %RH to 80 %RH 
(non-condensing) 

0 %RH to 100 %RH 
(non-condensing) 

5 %RH to 95 %RH 
(non-condensing) 

 

 Vibration  1.96 

m/s

2

 (10 Hz to 150 Hz)

1.96 m/s

2

 (10 Hz to 150 Hz)  9.8 m/s

2

 (10 Hz to 150 Hz) 

Color Pale 

gray 

Cover 

PC resin 

Materials 

Base 

Modified PPE resin 

Weight 

Approx. 80 g 

Installation 

With the mounting kit (sold separately). 
Lead wire (0.75 mm

2

 

×

 300 mm long) connection 

Connection 

Modular connection only for Model TY7043Z0P01(-1) 
 *  Modular jack is provided.  LAN cable compliant with EIA/TIA-568 Category 3 or over  

(

0.5 mm x 4 pairs) is required for modular connection. 

Accessory 

Four main unit mounting screws (M3 x 16) 

JIS: Japanese Industrial Standards 
PPE: Polyphenylene ether 

 
CE Marking Conformity 

Models TY70X3Z0P0X, HY7043 series and HTY7043 series comply with the following Electromagnetic Compatibility (EMC). 

 EMC: 

EN61326-1  Class B, Table 1 (For use in a basic electromagnetic environment) 

 

(CISPR11: 2003, EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-11) 

Note: 

Models TY7043Z0K00 and TY7043Z0K00-1 (Pt1000 temperature sensor model of Neosensor) are "equipment which is inherently benign in 
terms of electromagnetic compatibility" and "is excluded from the scope of the EMC Directive."  

Guide for the EMC Directive 2004/108/EC 

(8th February 2010) 

Содержание HTY7043T1P00

Страница 1: ...sing Neosensor is thus suitable for various applications such as commercial building air conditioning or other indoor uses Note The specification data of the two wire 4 20mA output model of Neosensor...

Страница 2: ...bligatory action to prevent possible danger The symbol inside graphically indicates the actual action to be carried out For example the sign on the left indicates general instructions CAUTION Use the...

Страница 3: ...ow Available model numbers Model number Type Power Humidity output Temperature output Signal output wire connection Company logo HTY7043T1P00 Yes HTY7043T1P00 1 Neosensor 24 V DC AC 1 5 VDC RTD Pt100...

Страница 4: ...iring Thermoplate for open wiring is used for open wiring installation Part No DY2000A3021 For mounting three Neosensor crosswise Auxiliary device Multi Thermocase Model TY1100Z LAN cable Part No DY72...

Страница 5: ...tion resistance Min 500 V DC 20 M Withstand voltage 500 V AC applied for 1 min with max of 1 mA leakage current between housing and terminals Environmental conditions Rated operating condition Limit o...

Страница 6: ...tead of lead wires Model TY7043Z0P01 1 Figure 1 Mounting dimensions mm Neosensor with Wall direct mounting kit Neosensor high sensitive type Note For wiring secure more than 15 mm clearance inside the...

Страница 7: ...unting surface when installing Neosensor with Thermoplate mounting kit 2 The number of lead wires differs depending on the models Note that the Pt100 temperature sensor model with modular connection h...

Страница 8: ...th Thermoplate mounting kit Figure 5 Mounting dimensions mm Neosensor high sensitive type with Thermoplate mounting kit on Thermoplate Thermoplate mounting kit Figure 6 Dimensions mm Mounting plate of...

Страница 9: ...sensor model with modular connection has a modular jack instead of lead wires Model TY7043Z0P01 1 Figure 7 Mounting dimensions mm Neosensor not covered with Multi Thermocase mounting kit Multi Thermoc...

Страница 10: ...rated by office device or equipment for example stays on Air circulation is interfered by furniture or door for example Temperature and or humidity sensing is affected by draft downdraft and hot cold...

Страница 11: ...14 5 Set the depth change levers on the bilateral sides of the Neosensor main unit to the upper position indicated with L shown in Fig 17 6 Mount the main unit of Neosensor on the main unit of the Wa...

Страница 12: ...d with L shown in Fig 17 6 Mount the main unit of Neosensor on the main unit of the Wall direct mounting kit with four mounting screws M3 16 supplied with Neosensor 7 Attach the cover back to the Neos...

Страница 13: ...on the bilateral sides of the Neosensor main unit to the upper position indicated with L shown in Fig 17 7 Mount the main unit of Neosensor on the main unit of the Thermoplate mounting kit with four...

Страница 14: ...lled with its cover removed IMPORTANT Before installing Neosensor in the Multi Thermocase be sure to set the depth change levers on the bilateral sides of the Neosensor main unit at the lower position...

Страница 15: ...pth change Raising the Neosensor main unit 3 Set the depth change levers located on the bilateral sides on the main unit front surface at the lower position The levers are factory set at the upper pos...

Страница 16: ...ck Blue Black 24 V AC DC 1 5 V DC 0 10 V DC Power supply Humidity Models HY7043T1000 1 and HY7043T6000 1 Figure 22 Terminal wiring Humidity sensor with 1 5V 0 10V humidity output 24 V AC DC power Pt10...

Страница 17: ...sor temperature humidity sensor models IMPORTANT Use isolated transformer to supply 24 V AC power Single Neosensor must be powered by a single 24 V AC transformer If the multiple Neosensor or a Neosen...

Страница 18: ...wing Use a controller with low pass filter 40 dB or higher removal ratio in normal mode that receives signals from Neosensor For insufficient removal ratio provide an isolator on the controller input...

Страница 19: ...ration Neosensor is factory calibrated and does not require field calibration right after installation However field calibrate Neosensor if a shift in the output due to aged deterioration is found For...

Страница 20: ...ademark of Azbil Corporation in Japan or in other countries Specifications are subject to change without notice Building Systems Company http www azbil com Rev 2 0 Sep 2015 AB 7050 J AI 7050 Rev 3 2 P...

Отзывы: