Delivering Current
The micro5 delivers the current either from standard electrodes, or in the Probe Mode using Q-Tips moist-
ened with tap water. We supply a sample of reusable self-adhesive electrodes. These types of electrodes
are readily available for sale through several places on the Internet, or through us.
When using Q-Tips, please purchase the paper shaft Q-Tip brand. There really is a difference. Never use
the plastic shaft (they don’t conduct electricity), and be sure to watch the instruction video on the proper
insertion of the Q-Tip.
Reusable electrodes have a life span of about 4-5 uses. The micro5 will alert you through it Secure Alert
warning tones if the electrode is failing.
If using reusable electrodes, make sure you prep the skin using alcohol swabs to remove as much skin oils
as possible. Oil does not conduct electricity AND it quickly breaks down the adhesive gel of the electrode
which is designed to both adhere to the skin and conduct electricity.