background image

17

More about the “I” Interferential Preset

Your AXION micro5 has a very special factory preset for microcurrent interferential. This setting is very 

useful in bathing a larger area with the stimulation. Below is a very simple ex planation of how interferntial 

works

The basic definition of interferential therapy is the use of two channels  of  stimulation crossing each other 

at 90°angles. At the point of intersection, an interference pattern is created (beat frequency) that is the dif-

ference of the frequencies of the two channels. For example: If channel one is 900 Hz and channel two is 

950Hz, the beat frequency created by the intersection of the two currents would be SO Hz (950-900). This 

third “beat” frequency is produced inside the body and radiates outward at  45° angles as shown in fig.1. 

This figure shows a line representation of “Static Interferential”.This representation holds  true if both  

channels  have the same microamp  output.

However if the output of the channels is varied, the current path of the beat frequency can be moved. This 

technique called “Vector Scan” is used to move the beat frequency path over a greater area. This is illus-

trated in fig.2.

Using the “I” Preset ch 1

.

When you press the “I” preset you will  notice that both chan-

nels are ready   for programming.

   

BOTH CHANNELS MUST BE USED. The microamp default for both chan-

nel is 400. 

Static Full Field with vector scan

The output of channel 1 is 900 Hz. This is fixed and cannot be changed. The 

default frequency of channel 2 is 150 Hz. This actually represents 1050 Hz. 

When the start key is pressed the frequency of channel two will begin to scan 

from 0-150. This is called “Rate Scan”.This represents 900-1050 Hz It is 

displayed as 0-150 to show you the beat frequency being created. If you want the beat frequency to be a 

certain number it can be changed, however if it is changed it will no longer scan.

You will also notice that during interferential operation, the microamp outputs of channels 1& 2 increase 

and decrease. This is Vector Scan and is used to create a full-field effect to bathe a larger area with the 

stimulation. The channels will vary about +/- 30% of the set value. For example, if 500 is the set value to 

Vector Scan, the value of channel two will scan down to about 350 while the value of channel  2 will scan 

up to about 650.

NOTE: Each time you want the Rate Scan Vector Scan option, you must press the “I’ preset. After stop-

ping, the channels will reset to a static mode. If you want to reset to the Rate Scan/Vector Scan, press  “I”.

Also note that after auto-shutoff in the “I” preset that the frequency of channel 2 resets to 000.

 

 

Содержание micro5

Страница 1: ...ESSOR CONTROLS ARE DESIGNED TO SIMPLIFY TREATMENT DELIVERY WE HAVE MADE THE OPERATION OF THE micro5 VERY SIMPLE IF YOU FOLLOW THE BASIC GUIDE LINES OF THE INSTRUCTION MANUAL REMEMBER MOST PROBLEMS ARE...

Страница 2: ...ssociated with high output milliamperes instruments Microcurrent stimulation is usually sub sensory and often imperceptible to the patient The current is able to pass through the skin with out much st...

Страница 3: ...essing this green key sets the parameters of the selected channels for the E default 4 M Pressing this yellow key sets the parameters of the selected channels for the M default 5 I Pressing this white...

Страница 4: ...the screen 12 START Pressing this key starts the countdown operation of the instrument It will also silence the beep that is heard every ten seconds during operation 13 STOP Pressing this key suspends...

Страница 5: ...ost effective applications available for microcur rent stimulation The following table illustrates the parameter setting for each of the modes These values will appear on the LCD screen when the mode...

Страница 6: ...when a large area is treated This preset uses both channels to deliver the current through four pads The micro5 uses both a rate scan and vector scan When using the I preset you will notice the numbe...

Страница 7: ...example if you wanted to enter 600 uA you would press 6 0 0 To enter 60uA you do not need to press 0 6 0 Simply press 6 0 If you make an entry error or change your mind you can press the CLR key to cl...

Страница 8: ...CLR to make your change You may program both channels at once by pressing the COMB combined key You may press the START key at any time during programming You do not have to be in the time setting to...

Страница 9: ...y is a difference Never use the plastic shaft they don t conduct electricity and be sure to watch the instruction video on the proper insertion of the Q Tip Reusable electrodes have a life span of abo...

Страница 10: ...metal sides of the probe See Video 4 Scan the area to be treated You will notice that when contact with the skin is made the micro5 emits a tone This tone rises and falls in pitch as the resistance i...

Страница 11: ...ecial salts compound It s clear that when using moistened q tips they should be soaked in a saline solution Doing that eliminates one variable you can control 1 2 Teaspoon of our mi cro5 Electrolyte s...

Страница 12: ...w tones indicates a successful initial start up test During Operation A series of descending tones along with a flashing battery indicator and flashing test indicator means a low battery charge You sh...

Страница 13: ...e a current test failure Note INSTANT ALERT current test operates by capturing samples to measure against what is entered on the screen At lower frequencies it may take the microprocessor a little ext...

Страница 14: ...or pain of central origin 3 T E N S devices should be used only under the continued supervision of a licensed practitioner 4 T E N S devices have no curative value 5 T E N S is a symptomatic treatment...

Страница 15: ...n in the continuous mode until NEXT 1 is pressed again or until the instrument is turned off by the On Off switch on the back of the unit Display Current NEXT 2 This feature is used to display the rel...

Страница 16: ...Sequential Treatment chaining The micro5 allows the clinician to link two treatments together to form a single continuous treatment This is sometimes called chaining To sequence treatments 1 Press the...

Страница 17: ...the I preset you will notice that both chan nels are ready for programming BOTH CHANNELS MUST BE USED The microamp default for both chan nel is 400 Static Full Field with vector scan The output of ch...

Страница 18: ...e the charging module in the wall when not in use Leaving it in the wall will shorten the life of the charging module The micro5 cannot be used while it is charging IMPORTANT NOTE It takes the micro5...

Страница 19: ...AXION micro5 We have tested and recommend the Anker Astro3E Our tests show that the Anker can recharge the micro5 from dead batteries to 90 charge in 1 hour NOTE It does not power the micro5 It recha...

Страница 20: ...el 1 The second powers Channel 2 The D cell battery pack is located inside the unit and is acessible by removing the four screws around the handle on the back of the unit and carefully sliding the tra...

Страница 21: ...ants the AXION micro5 to be free from defects for a period of two years from date of pur chase Factory further warrants all accessories including batteries and chargers to be free from defects for a p...

Отзывы: