2

Power Up

The micro5 is turned ON by a momentary switch located on the back of the instrument, in the lower Right 

hand corner. The switch will activate the unit with a slight touch and the entire LCD display will light up. 

The micros has a built in battery saver that will automatically turn the unit off within five minutes of idle 

time. When the micros is turned on again it will display the last treatment settings used. NOTE: The ON 

switch does NOT turn the unit off. The unit powers down automatically after five minutes of idle time.

When the instrument is first turned on, it will run an internal self-diagnostic test lasting about three sec

-

onds. During this test all the LCD segments will display. A successful test will be followed by a series 

of alternating high and low tones indicating the instrument is ready for operation. A failure of the test is 

followed by a series of tones along with a flashing TEST indicator on the LCD screen. If the micros should 

fail a power-up test, contact us for instructions. To facilitate return shipping in the event of a failure, we 

recommend keeping the original shipping container.

 

Introduction

Microcurrent instruments use very small amounts of current (micro-amperage) to treat many of the same 

conditions normally associated with high output (milliamperes) instruments. Microcurrent stimulation is 

usually sub-sensory and often imperceptible to the patient. The current is able to pass through the skin with-

out much stimulation to motor or sensory nerves. Therefore, microcurrent stimulation is comfortable to the 

patient and enhances patient compliance.

The AXION micro5 low volt, pulsed micro-amp stimulator utilizes micro-processor controls and touch pad 

operation. The ease of operation allows the clinician to quickly, accurately and efficiently administer the 

desired treatments to the patient. In addition, by using pre-programmed presets, the micro5 reduces confu-

sion of treatment selections often associated with microcurrent stimulators. A full manual over-ride is also 

offered for the clinician who wants to customize a treatment for a patient.

For the clinician, novice or experienced in microcurrent stimulation, the micro5 provides an easy to operate, 

yet highly sophisticated therapy system. The outstanding clinical results you will achieve with microcurrent 

stimulation will convince you of its place in a comprehensive pain management program.

The  AXION  micro5 is  the  most  advanced  microcurrent  stimulator  available.  Each time the micro5 is 

turned on it performs the following self-diagnostic tests.

RAM, ROM, CPU, Display, Output tests

During operation the micro5 uses distinct and different tones as well as display indicators to alert the opera-

tor of systems problems. 

Содержание micro5

Страница 1: ...ESSOR CONTROLS ARE DESIGNED TO SIMPLIFY TREATMENT DELIVERY WE HAVE MADE THE OPERATION OF THE micro5 VERY SIMPLE IF YOU FOLLOW THE BASIC GUIDE LINES OF THE INSTRUCTION MANUAL REMEMBER MOST PROBLEMS ARE...

Страница 2: ...ssociated with high output milliamperes instruments Microcurrent stimulation is usually sub sensory and often imperceptible to the patient The current is able to pass through the skin with out much st...

Страница 3: ...essing this green key sets the parameters of the selected channels for the E default 4 M Pressing this yellow key sets the parameters of the selected channels for the M default 5 I Pressing this white...

Страница 4: ...the screen 12 START Pressing this key starts the countdown operation of the instrument It will also silence the beep that is heard every ten seconds during operation 13 STOP Pressing this key suspends...

Страница 5: ...ost effective applications available for microcur rent stimulation The following table illustrates the parameter setting for each of the modes These values will appear on the LCD screen when the mode...

Страница 6: ...when a large area is treated This preset uses both channels to deliver the current through four pads The micro5 uses both a rate scan and vector scan When using the I preset you will notice the numbe...

Страница 7: ...example if you wanted to enter 600 uA you would press 6 0 0 To enter 60uA you do not need to press 0 6 0 Simply press 6 0 If you make an entry error or change your mind you can press the CLR key to cl...

Страница 8: ...CLR to make your change You may program both channels at once by pressing the COMB combined key You may press the START key at any time during programming You do not have to be in the time setting to...

Страница 9: ...y is a difference Never use the plastic shaft they don t conduct electricity and be sure to watch the instruction video on the proper insertion of the Q Tip Reusable electrodes have a life span of abo...

Страница 10: ...metal sides of the probe See Video 4 Scan the area to be treated You will notice that when contact with the skin is made the micro5 emits a tone This tone rises and falls in pitch as the resistance i...

Страница 11: ...ecial salts compound It s clear that when using moistened q tips they should be soaked in a saline solution Doing that eliminates one variable you can control 1 2 Teaspoon of our mi cro5 Electrolyte s...

Страница 12: ...w tones indicates a successful initial start up test During Operation A series of descending tones along with a flashing battery indicator and flashing test indicator means a low battery charge You sh...

Страница 13: ...e a current test failure Note INSTANT ALERT current test operates by capturing samples to measure against what is entered on the screen At lower frequencies it may take the microprocessor a little ext...

Страница 14: ...or pain of central origin 3 T E N S devices should be used only under the continued supervision of a licensed practitioner 4 T E N S devices have no curative value 5 T E N S is a symptomatic treatment...

Страница 15: ...n in the continuous mode until NEXT 1 is pressed again or until the instrument is turned off by the On Off switch on the back of the unit Display Current NEXT 2 This feature is used to display the rel...

Страница 16: ...Sequential Treatment chaining The micro5 allows the clinician to link two treatments together to form a single continuous treatment This is sometimes called chaining To sequence treatments 1 Press the...

Страница 17: ...the I preset you will notice that both chan nels are ready for programming BOTH CHANNELS MUST BE USED The microamp default for both chan nel is 400 Static Full Field with vector scan The output of ch...

Страница 18: ...e the charging module in the wall when not in use Leaving it in the wall will shorten the life of the charging module The micro5 cannot be used while it is charging IMPORTANT NOTE It takes the micro5...

Страница 19: ...AXION micro5 We have tested and recommend the Anker Astro3E Our tests show that the Anker can recharge the micro5 from dead batteries to 90 charge in 1 hour NOTE It does not power the micro5 It recha...

Страница 20: ...el 1 The second powers Channel 2 The D cell battery pack is located inside the unit and is acessible by removing the four screws around the handle on the back of the unit and carefully sliding the tra...

Страница 21: ...ants the AXION micro5 to be free from defects for a period of two years from date of pur chase Factory further warrants all accessories including batteries and chargers to be free from defects for a p...

Отзывы: