The AXION micro5 comes with two probes. (Active and Indifferent). 

The probe with the button is the active probe used during treatment 

and conductivity measurements. The probe is handy to treat irregular 

body areas.

The black switch on the probe is a momentary one. You need only to 

touch it once. Pressing this button is the same as pressing the START 

key. In fact, should the button ever fail to begin the timer countdown 

you may press the START key to over-ride the probe start button.

CAUTION: REMOVE THE PROBE AFTER USE. LEAVING THE PROBE PLUGGED IN WILL 

CAUSE A FAILURE TEST TO APPEAR IN CHANNEL ONE IF YOU ARE TRYING TO USE THE 

PADS IN CHANNEL ONE.

Using the Probe

1. 

Press the PROBE key

2. 

Have the patient hold the metal part of the Indifferent probe in the hand to be treated.

3. 

Place a well moistened cotton swab tip into the end of the probe shaft The swab tip should be  broken  

 

off just below where the cotton attaches to the shaft. Use only paper shafted swabs. The cotton tip

  

delivers the current. Make sure it is both wet and making good contact with the metal sides of the 

 

probe. See Video

4. 

Scan the area to be treated. You will notice that when contact with the skin is made the micro5 emits a  

 

tone. This tone rises and falls in pitch as the resistance increases and decreases. The higher the pitch

  

the lower the resistance. You will notice the numbers on the LCD display will rise and fall. You are

 

 looking for the highest pitch in the area to be treated.

5. 

Press the black button on the probe to begin treatment. The seconds or minute indicator in channel one

 

 will begin to flash to signal timer countdown.

6. 

To leave the probe mode press the PROBE key.

When using the probes, both active and indifferent must be in contact with the body at the same time. Either 

both wet Q-Tips are used, or the active probe is used while the patient holds the metal part of the indifferent 

probe in their hand, usually on the same side of the body as is being treated.

Remember to remove the probe from the unit to use pads in channel 1.

Use one cup of tap water with 1/2 teaspoon of table salt added. This creates a normal sa-

line solution. Moisten the Q-Tips before inserting into the probes, and then for 10 seconds 

after inserting. Re-moisten after treating 2-3 points, or each muscle.

 

Using the Probe

Indifferent 

Probe

Active Probe

Содержание micro5

Страница 1: ...ESSOR CONTROLS ARE DESIGNED TO SIMPLIFY TREATMENT DELIVERY WE HAVE MADE THE OPERATION OF THE micro5 VERY SIMPLE IF YOU FOLLOW THE BASIC GUIDE LINES OF THE INSTRUCTION MANUAL REMEMBER MOST PROBLEMS ARE...

Страница 2: ...ssociated with high output milliamperes instruments Microcurrent stimulation is usually sub sensory and often imperceptible to the patient The current is able to pass through the skin with out much st...

Страница 3: ...essing this green key sets the parameters of the selected channels for the E default 4 M Pressing this yellow key sets the parameters of the selected channels for the M default 5 I Pressing this white...

Страница 4: ...the screen 12 START Pressing this key starts the countdown operation of the instrument It will also silence the beep that is heard every ten seconds during operation 13 STOP Pressing this key suspends...

Страница 5: ...ost effective applications available for microcur rent stimulation The following table illustrates the parameter setting for each of the modes These values will appear on the LCD screen when the mode...

Страница 6: ...when a large area is treated This preset uses both channels to deliver the current through four pads The micro5 uses both a rate scan and vector scan When using the I preset you will notice the numbe...

Страница 7: ...example if you wanted to enter 600 uA you would press 6 0 0 To enter 60uA you do not need to press 0 6 0 Simply press 6 0 If you make an entry error or change your mind you can press the CLR key to cl...

Страница 8: ...CLR to make your change You may program both channels at once by pressing the COMB combined key You may press the START key at any time during programming You do not have to be in the time setting to...

Страница 9: ...y is a difference Never use the plastic shaft they don t conduct electricity and be sure to watch the instruction video on the proper insertion of the Q Tip Reusable electrodes have a life span of abo...

Страница 10: ...metal sides of the probe See Video 4 Scan the area to be treated You will notice that when contact with the skin is made the micro5 emits a tone This tone rises and falls in pitch as the resistance i...

Страница 11: ...ecial salts compound It s clear that when using moistened q tips they should be soaked in a saline solution Doing that eliminates one variable you can control 1 2 Teaspoon of our mi cro5 Electrolyte s...

Страница 12: ...w tones indicates a successful initial start up test During Operation A series of descending tones along with a flashing battery indicator and flashing test indicator means a low battery charge You sh...

Страница 13: ...e a current test failure Note INSTANT ALERT current test operates by capturing samples to measure against what is entered on the screen At lower frequencies it may take the microprocessor a little ext...

Страница 14: ...or pain of central origin 3 T E N S devices should be used only under the continued supervision of a licensed practitioner 4 T E N S devices have no curative value 5 T E N S is a symptomatic treatment...

Страница 15: ...n in the continuous mode until NEXT 1 is pressed again or until the instrument is turned off by the On Off switch on the back of the unit Display Current NEXT 2 This feature is used to display the rel...

Страница 16: ...Sequential Treatment chaining The micro5 allows the clinician to link two treatments together to form a single continuous treatment This is sometimes called chaining To sequence treatments 1 Press the...

Страница 17: ...the I preset you will notice that both chan nels are ready for programming BOTH CHANNELS MUST BE USED The microamp default for both chan nel is 400 Static Full Field with vector scan The output of ch...

Страница 18: ...e the charging module in the wall when not in use Leaving it in the wall will shorten the life of the charging module The micro5 cannot be used while it is charging IMPORTANT NOTE It takes the micro5...

Страница 19: ...AXION micro5 We have tested and recommend the Anker Astro3E Our tests show that the Anker can recharge the micro5 from dead batteries to 90 charge in 1 hour NOTE It does not power the micro5 It recha...

Страница 20: ...el 1 The second powers Channel 2 The D cell battery pack is located inside the unit and is acessible by removing the four screws around the handle on the back of the unit and carefully sliding the tra...

Страница 21: ...ants the AXION micro5 to be free from defects for a period of two years from date of pur chase Factory further warrants all accessories including batteries and chargers to be free from defects for a p...

Отзывы: