- 15 -
You will see a window with two edit boxes for minimum and
maximum reflectance values.
Ensure that the values shown in the window match the values printed
at the bottom of the calibration sheet.
(Modify the figures on the screen if necessary to ensure that they are
the same as the card
With the Calibration window open scan the calibration barcode 10
times. Then click the OK button.
The verifier is now calibrated and ready to use.
After calibration is completed, store the calibration sheet in its wallet,
away from light and physical damage. If your calibration sheet should
ever become damaged or discoloured in any way, please contact Axicon
or your supplier for a replacement.
NB: During the calibration process the software displays the
uncalibrated data and therefore reflectance values and grades
displayed
behind
the calibration window can be ignored.
Under normal operating conditions Axicon recommends calibrating the
verifier monthly (the verifier software is set up to remind you when it is
time to calibrate). In variable light and temperature conditions we
recommend more frequent calibration. The frequency can be changed if
required by using the “Options / Set-up /Calibration” path.