- 14 -
USER CALIBRATION
Calibration with known reflectance values is essential to ensure the
accuracy of this equipment. It should be carried out after installation and
at regular intervals. (see below)
When the software detects an uncalibrated reader the following dialogue
box will appear “Reader XXXX has not been calibrated”. You must click
OK to continue.
Please switch on your verifier at least five minutes before you start
to calibrate.
To calibrate you will need your calibration sheet, which you will find
in the wallet.
Take the calibration sheet out of the wallet and place it on a flat
surface.
Check that it has not been damaged in any way
For the Axicon 6000 Series use the EAN13 calibration code at the top
left of the sheet.
For the Axicon 6500 and 7000 Series use either of the large
calibration codes at the bottom of the sheet.
Select the calibrate option.
This is located in the “Options” menu….
(or you can use the icon)