7 — Calibration and Adjustments
7–4
CX20/CX30/CX34 Service Manual
7.5
Calibration
To calibrate the unit, proceed as follows:
NOTE:
See Section 5.2 for service mode entry.
What You Do
What You See
1
On the CX20/CX30/CX9/A702 check
that unit is level and adjust levelling
feet if necessary.
2
Enter service mode by using the DIL
switch (refer to Section 5.2.2).
CONFIGURATION
3
Press
→
(or
←
) to obtain the
calibration menu option.
CALIBRATION
4
Press ENTER.
10
EMPTY PAN
3 : 6
3/4
5
Ensure weighplate is empty, then
press ENTER.
7
ADD FULL LOAD
0 : 0
6
Place full load on the weighplate.
4
ADD FULL LOAD
19 : 2
3/8
7
Press ENTER.
10
EMPTY PAN
29 : 6
3/8
8
Remove load.
9
EMPTY PAN
4 : 6
1/2
9
Press ENTER.
7
ADD FULL LOAD
0 : 0
Содержание CX20 SERIES
Страница 133: ...11 Main PCB 11 4 CX20 CX30 CX34 Service Manual Figure 11 1 Main PCB Electrical Connections See Tables 11 1 and 11 2 ...
Страница 137: ...11 Main PCB 11 8 CX20 CX30 CX34 Service Manual Figure 11 2 Main PCB DIL Switches and Potentiometers ...
Страница 246: ...25 Appendices 25 9 CX20 CX30 CX34 Service Manual 1 Voids printed on receipt 2 Voids not printed on receipt ...
Страница 258: ...PL 1 CX20 CX30 CX34 Service Manual Section PL Parts List ...
Страница 284: ......
Страница 287: ......
Страница 317: ...Addendum A Phase 1 4 and CX34 Parts List A 30 CX20 CX30 CX34 Addendum A ...
Страница 318: ......
Страница 321: ......
Страница 323: ...Contents B 2 Addendum B ...
Страница 343: ...Addendum B Phase 1 5 1 5 3 and New Main PCB B 22 Addendum B ...
Страница 344: ......
Страница 347: ......
Страница 351: ...PL Parts List Devlin Keyboard C 6 Addendum C ...
Страница 352: ......