H EMC
Warning
◆
The devices need to be installed and used in compliance with the electromagnetic
environment.
◆
Portable and mobile RF communications equipment can affect the equipment
◆
Pins of connectors identified with the ESD warning symbol should not be touched and
connections should not be made to these connectors unless ESD precautionary procedures
are used.
◆
The use of accessories, sensor and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the equipment as replacement parts for
internal components, may result in increased emissions or decreased immunity of the
equipment.
◆
The equipment should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the equipment should be observed to verify normal
operation in the configuration in which it will be used.
The equipment cannot be used magnetic resonance (MRI) environment.
Guidance an
d manufacture’s declaration – electromagnetic emission
The Anesthetic Workstation is intended for use in the electromagnetic environment specified
below. The customer or the user of ANESTHETIC MACHINE should assure that it is used in such
an environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The Anesthetic Workstation uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emission
CISPR 11
Class A
The Anesthetic Workstations suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacture’s declaration – electromagnetic immunity
The Anesthetic Workstation intended for use in the electromagnetic environment specified below. The
customer or the user of ANESTHESIA MACHINE should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
6 kV contact
6 kV contact
Floors should be wood, concrete