6.7.8.3 O
2
Cell Calibration
NOTE:
⚫
The O
2
cell calibration requires that an O
2
cell be installed, monitoring enabled, and oxygen be
available.
1. Touch [
O
2
cell
] button, the calibration menu appears.
2. Follow the instructions displayed, disconnect the O
2
sensor from breathing circuit, touch
[
100%
] button or [
21%
] button respectively.
3. When calibration is complete, verify that the message bar displays
“
Calibration
Completed!
”.
6.7.9 Service Modes Menu
Touch the [
Service Modes]
window (
Figure 6-34
) to view and service.
Service is only accessible
when the correct password has been given.
Figure 6-34
Service Modes Menu