28
Atmospheric Technology
converters in the embedded computer do not read negative voltages. Therefore, to
detect MFC negative zero drift, a zero bias needs to be applied to the MFC. Once
the zero is established, the slope is applied to adjust the MFC span to agree with a
certified flow standard referenced to EPA standard conditions of 25
°
C and 760mm
Hg.
It is important to understand that the flow values shown on the Manual Screen can
be referenced to two conditions, either the MFC manufacturer’s reference condition
of 0
°
C and 760 mmHg, or another referenced condition, usually 25
°
C and 760
mmHg. The latter is the condition used during sampling and also displayed in red on
the Manual Screen when the “Calibrated Flow” box is checked.
When it becomes necessary to re-calibrate the MFCs, the flowing procedure should
be followed:
1. Go to the Advanced screen and change the slope for the MFC to be
calibrated to 1.00 and change the intercept value to the negative value of
1.5% of the full MFC range (the MFC range is displayed in the Advanced
Screen). For example, if the MFC has a full range of 2.0 lpm, the intercept
value should be set to -0.030.
2. Go to the Manual screen and check the box next to “Calibrated Values”. With
no flow, mechanically adjust the MFC zero potentiometer until the screen
display for the flow rate is close to zero (refer to the MFC Operations Manual
in the Appendix for adjustment procedures). There will be some small
oscillations around zero when doing this but try to get it as close as possible.
3. Activate the necessary valve(s) and pump to establish flow rate through the
MFC. Set the MFC flow set value to the nominal flow (the flow rate most
commonly used in sampling). Double press the Set button to apply the new
flow rate set value. If there is no Set button then pressing the Accept button
will set the flow rate.
Содержание 8000
Страница 2: ......
Страница 4: ...ii Atmospheric Technology ...
Страница 8: ...2 Atmospheric Technology ...
Страница 14: ...8 Atmospheric Technology ...
Страница 18: ...12 Atmospheric Technology ...
Страница 30: ...24 Atmospheric Technology ...
Страница 39: ...Atmospheric Technology 33 Appendix A Schmatics ...
Страница 40: ...34 Atmospheric Technology ...
Страница 41: ...1 Model 2200V11 Interface Board Schematic 1 sch 7 ...
Страница 42: ......
Страница 43: ...Model 2200V11 Interface Board Schematic 2 sch 2 7 ...
Страница 44: ......
Страница 45: ...Model 2200V11 Interface Board Schematic 3 sch 3 7 ...
Страница 46: ......
Страница 47: ...Model 2200V11 Interface Board Schmatic 4 sch 4 7 ...
Страница 48: ......
Страница 49: ...Model 2200V11 Interface Board Schematic 5 sch 5 7 ...
Страница 50: ......
Страница 51: ...Model 2200V11 Interface Board Schematic 6 sch 6 7 ...
Страница 52: ......
Страница 53: ...Model 2200V11 Interface Board Schematic 7 sch 7 7 ...
Страница 54: ......
Страница 55: ...Atmospheric Technology 49 Appendix B Manual for Mass Flow Controller ...
Страница 56: ......