14
Atmospheric Technology
Ch.2 MFC Set Point
The sampling flow rate for the Channel 2 cartridge is specified in this box
Flow Leak Limit
The flow leak limit is the maximum flow rate that is allowed for a cartridge to pass
either manual or automatic leak check. This value has the same units as the flow
rate.
Flow Tolerance
The cartridge flow rate is continuously monitored during sampling. If the flow rate
specified in Ch.1 MFC or Ch.2 MFC Set Point box is outside this tolerance, an error
status for flow tolerance will be generated. This value has the same units as the
flow rate.
Cycled Schedule
The Cycled Schedule value specifies the time interval in days between schedules.
This allows the operator to schedule repeated samples at a set interval automatically
(i.e. sampling repeated every 6 days if set to 6). The value must be greater than 1 to
be applied.
Purge Time
Purge time specifies the purging time in minutes of the inlet line and inlet manifold
before sampling the cartridges.
Data Write Interval
The Data Write Interval specifies the time interval for recording canister pressure
and cartridge flow rate data. The minimum time interval is 5 minutes.
Site Label
Содержание 8000
Страница 2: ......
Страница 4: ...ii Atmospheric Technology ...
Страница 8: ...2 Atmospheric Technology ...
Страница 14: ...8 Atmospheric Technology ...
Страница 18: ...12 Atmospheric Technology ...
Страница 30: ...24 Atmospheric Technology ...
Страница 39: ...Atmospheric Technology 33 Appendix A Schmatics ...
Страница 40: ...34 Atmospheric Technology ...
Страница 41: ...1 Model 2200V11 Interface Board Schematic 1 sch 7 ...
Страница 42: ......
Страница 43: ...Model 2200V11 Interface Board Schematic 2 sch 2 7 ...
Страница 44: ......
Страница 45: ...Model 2200V11 Interface Board Schematic 3 sch 3 7 ...
Страница 46: ......
Страница 47: ...Model 2200V11 Interface Board Schmatic 4 sch 4 7 ...
Страница 48: ......
Страница 49: ...Model 2200V11 Interface Board Schematic 5 sch 5 7 ...
Страница 50: ......
Страница 51: ...Model 2200V11 Interface Board Schematic 6 sch 6 7 ...
Страница 52: ......
Страница 53: ...Model 2200V11 Interface Board Schematic 7 sch 7 7 ...
Страница 54: ......
Страница 55: ...Atmospheric Technology 49 Appendix B Manual for Mass Flow Controller ...
Страница 56: ......