19. Specifications
Cat.No.
3564
Measurement scale
Brix(%) (Automatic Temperature Compensation), Temperature
Measurement range
Brix 0.00 to 80.0% (indication: -2.0 to 80.5%),
Temperature 5 to 100℃ (indication: 0 to 130℃)
Resolution
Brix 0.01 or 0.1% (by selection; 0.00 to 9.99%)
Brix 0.1% (10.0 to 80.0%)
Measurement accuracy
Brix ±0.1% (based on sucrose solution)
Measurement temperature
5.0 to 100.0℃ (Automatic Temperature Compensation range)
Temperature correction
values
Based on the temperature correction table for sucrose.
For 15 to 40℃, ICUMSA (International Committee of Unified Method of Sugar
Analysis) values are used.
For 5 to 15℃ or 40 to 100℃, the values obtained at the ATAGO laboratories are
used.
The temperature correction values for sucrose can be multiplied by any factor in
the range of 0.80 to 5.00.
Output (Optional Cables)
①Recorder output: DC 4 to 20mA
Any value between Brix 1% through 80% can be set.
②RS-232C output
Source
LED (D-line approximation)
Temperature sensor
Thin film platinum sensor
Materials in contact with the
solution
Prism
: Artificial sapphire
Prism stage : SUS316
Wetted parts materials
-30 to 150℃
Resistible pressure on the
prism unit
0.98MPa(10kgf/cm
2
)
International Protection
Class
IP64
Power supply
100 to 240V (50/60Hz) (voltage fluctuation not to exceed ±10%)
CM-800α: DC24V (Allowable fluctuation is ±10%)
AC adapter AD-32, AD-33 or AD-34 (optional): AC 100-240V 50/60Hz
Power consumption
3VA
Environmental conditions
・ Use the instrument at an altitude below 2,000m (above sea level).
・ Use the instrument indoors.
・ Use the instrument where the temperature is between 5 to 40℃.
・ Use the instrument under the condition where humidity is 80% at 31℃ or lower,
falling linearly to 50% at 40℃.
・ Main supply voltage fluctuation should not to exceed ±10% the nominal voltage.
・ Installation categories (Overvoltage Categories) : ll
・ The pollution degree is 2 (according to IEC60664).