-26-
13. Setting the Temperature Correction Values
●About temperature correction
The Refractive Index of every substance varies with the temperature. When measuring the Refractive
Index of a liquid with a refractometer, the measurement value will also vary according to the sample
temperature.
The CM-800α detects the temperature of the prism, and the Brix(%) indication values are automatically
corrected based upon the preset temperature correction values. When the sample temperature is within
the range of 5 to 100℃, the Brix(%) values are automatically compensated for temperature.
●Setting the temperature correction values for the CM-800α
When the CM-800α is shipped from the factory, the temperature correction values for sucrose have
been set as the default. (C Table 13-1 page 27)
Fig.13-1 lists the temperature correction values for temperatures between 15 to 40℃.
However, temperature correction values for 5 to 100℃ have been set on the CM-800α.
If sugar is the main constituent of soluble solids in the sample, these values can be used as they are.
If the temperature correction factor is changed inadvertently, refer to page 28 "Inputting Temperature
Correction Factor". By utilizing these instructions, it is possible to set the temperature correction factor
back to "1.00".
●Setting procedure
If the main constituent of the sample is not sucrose, a different temperature correction value must be set.
To set a new temperature correction value, you must determine the relationship between the temperature
correction factor for sucrose and your desired sample. The temperature correction factor for sucrose is
1.00. When a new temperature correction factor is input, this factor is multiplied by the temperature
correction factor for sucrose.
The temperature correction factor can be set in the range of 0.80 to 5.00.
EXAMPLE :
If the Refractive Index variation of a sample due to temperature is 3 times larger than that of sucrose, the
temperature correction factor should be set to 3.00.