Astralpool 54568 Скачать руководство пользователя страница 2

User calibration

: cAL

Manufacturing calibration : CAL

Zero

Test

Mode

On

Off

CAL

SEL

SEL

CAL

Mode

On

Off

cAL

SEL

Calibration Mode

On

Off

Mode

CAL

Cl

CAL

R E S U L T   +   x

R E S U L T   +   x

Mode

Zero

Test

0.0.0

Zero

Test

R E S U L T

R E S U L T

CAL

Zero

Test

On

Off

:             :

Note

CAL

cAL

M E T H O D

M E T H O D

M E T H O D

M E T H O D

CAL

Recommended calibration values

Chlorine:

between 0,5 and 1,5 mg/l*

pH:

between 7,6 and 8,0*

Cyanuric acid:

between 30 and 60 mg/l

Total Alkalinity:

between 50 and 150 mg/l CaCO

3

Calcium Hardness: between 100 and 200 mg/l CaCO

3

* or rather values mentioned in the reference standard kits

User notes

E 10

E 70

E 72

E 74

E 71

E 73

E 75

Calibration factor "out of range"

Cl:

Manufacturing calibration incorrect / erase

pH:

Manufacturing calibration incorrect / erase

Cys:

Manufacturing calibration incorrect / erase

tA:

Manufacturing calibration incorrect / erase

CA.H: Manufacturing calibration incorrect / erase

Cl:

User calibration incorrect / erase

pH:

User calibration incorrect / erase

Cys:

User calibration incorrect / erase

tA:

User calibration incorrect / erase

CA.H: User calibration incorrect / erase

Press MODE key and 

keep it depressed

.

Switch unit on using ON/OFF key.
Release MODE key after approx. 1 second.

Select the test required using the MODE key:
CAL

 

Cl 

→ 

CAL pH 

→ 

CAL Cys 

→ 

CAL tA 

→ 

CAL CA.H .

 

... (Scroll)

Perform zero calibration (see "Operation").
Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The display shows the following in alternating mode:

Place the calibration standard to be used in the sample
chamber with the 

Δ

 and 

 marks aligned.

Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The result is shown in the display, alternating with CAL.

If the result displayed corresponds with the value of the
calibration standard (within the tolerance quoted), exit
calibration mode by pressing the ON/OFF key.

Otherwise, pressing the MODE key once increases the
displayed value by 1 digit.
Pressing the ZERO/TEST key once decreases the
displayed value by 1 digit.

Pressing the relevant key until the displayed value equals
the value of the calibration standard.

By pressing the ON/OFF key, the new correction factor
is calculated and stored in the user calibration software.

Confirmation of calibration (3 seconds).

Factory calibration active.

Calibration has been set by the user.

To reset the calibration to the factory setting:

Press both the MODE and ZERO/TEST and 

keep them

depressed

.

Switch the unit on using the ON/OFF key. Release the
MODE and ZERO/TEST keys after approx. 1 second.

The following messages will appear in turn on the display:

The calibration is reset to the factory setting.
(SEL stands for Select)

or:

Calibration has been set  by the user. (If the user calibration
is to be retained, switch the unit off using the ON/OFF
key.)

Calibration is reset to the factory setting by pressing the
MODE key. The following messages will appear in turn
on the display:

Switch the unit off using the ON/OFF key.

Technical changes without notice
Printed in Germany 07/98

E 76

E 77

1. Vials, caps and stirring rods should be cleaned thoroughly 

after each

analysis

 to prevent errors being carried over. Even minor reagent

residues can cause errors in the test results. Use the brush provided for
cleaning.

2. The outside of the vial must be clean and dry before starting the analysis.

Clean the outside of the vials with a towel. Fingerprints or other marks
will be removed.

3. Zero calibration and test must be carried out with the same vial as there

may be slight differences in optical performance between vials.

4. The vials must be positioned in the sample chamber for zero calibration

and test with the 

Δ

-mark on the vial aligned with the 

∇−

mark on the

instrument.

5. Place the cover on the sample chamber for zero calibration and test.

6. Bubbles on the inside of the vial may also lead to errors. In this case, fit

the vial with a clean stopper and remove bubbles by swirling the contents
before starting test.

7. Avoid spillage of water in the sample chamber. If water should leak into

the photometer housing, it can damage electronic components and
cause corrosion.

8. Contamination of the windows over the light source and photo sensor

in the sample chamber can result in errors. If this is suspected check
the condition of the windows.

9. When using reagent tablets, use only tablets in black printed foil. For pH

value determination, the PHENOLRED-tablet foil should also be marked
PHOTOMETER.

10.The reagent tablets should be added to the water sample without being

handled.

11.Large temperature differentials between the photometer and the

operating environment can lead to incorrect measurement due to the
formation of condensate in the area of the lens or on the vial (e.g).

12. To avoid errors caused by stray-light do not use the instrument in bright

sunlight.

Troubleshooting: Guidelines for photometric measurements

E 78

E 79

Correct filling of the vial

correct

wrong

10 ml

10 ml

Содержание 54568

Страница 1: ... the cap tightly and place the vial in the sample chamber making sure the Δ and marks are aligned Press the ZERO TEST key The method symbol flashes for approx 3 seconds The mg l CaCO3 value is shown in the display Rinse the vial and cap thoroughly after each test Tolerance 5 Full Scale Total Alkalinity 5 200 mg l CaCO3 0 0 0 R E S U L T R E S U L T Zero Test tA Switch the unit on using the ON OFF ...

Страница 2: ...ess both the MODE and ZERO TEST and keep them depressed Switch the unit on using the ON OFF key Release the MODE and ZERO TEST keys after approx 1 second The following messages will appear in turn on the display The calibration is reset to the factory setting SEL stands for Select or Calibration has been set by the user If the user calibration is to be retained switch the unit off using the ON OFF...

Страница 3: ...ondes Les résultat s affiche en mg l chlore total c Chlore combin Chlore combiné chlore total chlore libre Tol rance 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Procéder au calibrage du zéro voir Mise en service Ajouter une pastille de PHENOLRED PHOTOMETER dans l échantillon de 10 ml en la sortant directement de la feuille de protection sans la to...

Страница 4: ...de CAL Cl CAL RESULTAT x RESULTAT x Mode Zero Test 0 0 0 Zero Test R E S U L T A T R E S U L T A T CAL Zero Test On Off Remarque CAL cAL M E T H O D E M E T H O D E M E T H O D E M E T H O D E Valeures de calibrage recommand Chlore entre 0 5 et 1 5 mg l pH entre 7 6 et 8 0 Acide cyanurique entre 30 et 60 mg l Alcalinité totale entre 50 et 150 mg l CaCO3 Dureté calcique entre 100 et 200 mg l CaCO3 ...

Страница 5: ... Tolerancias 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Realizar la calibraci n a cero v ase instrucciones A los 10 ml de prueba acuosa a adir directamente de su envoltura una tableta PHENOLRED PHOTOMETER y machacarla con una varilla limpia Disolver completamente la tableta cerrar la cubeta con su tapa y posicionarla Δ Presionar ZERO TEST El s mb...

Страница 6: ...roduce errores de medici n En este caso se tapa la cubeta y las burbujas se disuelven girando la cubeta antes de realizar el an lisis 7 La infiltraci n de agua en la c mara de medici n debe ser evitada La entrada de agua en la caja del fot metro puede destruir las piezas de construcci n electr nicas y producir da os de corrosion 8 El ensuciamiento de la ptica diodo luminoso y fotosensor en la c ma...

Страница 7: ... de método presionar la tecla MODE CAL Cl CAL pH CAL Cys CAL S 4 3 Scroll Realizar la calibración a cero como descrito Presionar la tecla ZERO TEST El símbolo del método parpadea durante aproximadamente 3 segundos En el display aparece alternativamente Colocar el estándar a utilizar en el compartimento de medición según las marcas Presionar la tecla ZERO TEST El símbolo del método parpadea durante...

Страница 8: ...y appare il risultato in mg l del cloro totale c Cloro combinato Cloro legato cloro totale cloro libero Tolleranza 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Condurre la taratura a zero vedere la messa in funzione Aggiungere direttamente nei 10 ml dell acqua di prova una compressa di PHENOLRED PHOTOMETER senza toccarla con le dita e frantumarla c...

Страница 9: ... e le bollicine si dissolvono agitando la cuvetta prima di procedere al test 7 Si deve impedire la penetrazione di acqua nel pozzetto di misurazione L entrata dell acqua nel corpo del fotometro pu comportare un danneggiamento dell elettronica e danni dovuti a corrosione 8 L apparecchiatura ottica diodo luminoso e fotosensore nel pozzetto di misurazione se sporca comporta errori di misurazione Le s...

Страница 10: ...ffettuata dall operatore scorretto cancellato E 10 E 70 E 72 E 74 E 71 E 73 E 75 E 76 E 77 E 79 E 78 Calibrazione effettuata dall operatore cAL Calibrazione effettuata in fabbrica CAL Zero Test Mode On Off CAL SEL SEL CAL Mode On Off cAL SEL Indicazione Metodo di calibrazione On Off Mode CAL Cl CAL RISULTATO x RISULTATO x Mode Zero Test 0 0 0 Zero Test R I S U L T A T O R I S U L T A T O CAL Zero ...

Страница 11: ...rdar 2 minutos por uma reacção colorida Pressione a tecla ZERO TEST O símbolo CA H piscará durante 3 segundos O resultado aparecerá no visor expresso em mg l CaCO3 Tolerância 5 em toda a escala Efectue a leitura do Zero Ver manuseamento Retire a célula do compartimento de teste adicione um comprimido ALKA M PHOTOMETER e esmague o até se dissolver por completo usando um triturador limpo Feche a cél...

Страница 12: ...itura Se houver suspeita de contaminação examine o visor 9 Quando usar reagentes em comprimidos use apenas comprimidos protegidos com folha metálica No caso dos comprimidos PHENOLRED devem estar marcados na folha metálica com PHOTOMETER 10 Os comprimidos devem ser adicionados à amostra de água sem serem tocados com os dedos 11 Elevadas diferenças de temperatura entre o fotómetro e o ambiente de tr...

Страница 13: ...n de display verschijnt het meetresultaat in mg l totaal chloor c gebonden Chloor gebonden chloor totaal chloor vrij chloor Toleranties 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Doe de nulstelling zie In bedrijf nemen Doe een PHENOLRED PHOTOMETER tablet direkt uit de folie in 10 ml monsterwater en maak deze fijn met een schone roerstaaf Tablet g...

Страница 14: ...igingen voorbehouden Printed in Germany 09 00 Druk op de MODE toets en houd deze ingedrukt Schakel de fotometer in met de ON OFF toets laat na ca 1 seconde de MODE toets los Druk voor het wisselen van het meetbereik de MODE toets CALCl CALpH CALCys CAL tA CALCA H Scroll Doe de normale nulstelling Druk de ZERO TEST toets Het methode symbool knippert ca 3 seconden In de aanwijzing verschijnt afwisse...

Страница 15: ...EST Metodesymbolet blinker i ca 3 sekunder Displayet viser resultatet i mg l klor total c bundet klor bundet klor klor total frit klor Tolerance 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Gennemfør nulkompensation se ibrugtagning En PHE NOLRED PHOTOMETER tablet tilsættes til 10 ml vandprøven direkte fra folien og knuses med en ren rørepind Opløs ...

Страница 16: ...tiveres fabrikations kalibreringen Displayet viser skiftevis Apparatet slukkes på tasten ON OFF Kalibrering out of range Cl Fabrikations kalibrering ukorrekt slettet pH Fabrikations kalibrering ukorrekt slettet Cys Fabrikations kalibrering ukorrekt slettet TA Fabrikations kalibrering ukorrekt slettet CAH Fabrikations kalibrering ukorrekt slettet Cl Bruger kalibrering ukorrekt slettet pH Bruger kal...

Отзывы: