Astralpool 54568 Скачать руководство пользователя страница 11

Fonte de luz:

2 LED: 

λ

1

 = 528 nm (filtro) ; 

λ

2

= 605 nm

Bateria:

9 V (Vida útil 600 testes).

Auto-OFF:

5 minutos depois de ter premido pela última
vez uma tecla.

Condições ambientais:

5-40

°

C

humidade relativa (sem condensação)

CE:

DIN EN 55 022, 61 000-4-2, 61 000-4-8,
50 082-2, 50 081-1, DIN V ENV 50 140, 50 204

Dados técnicos

Manuseamento

5 in 1 Fotómetero :  Cl,  pH,  Cys,  TA,  CAH

Mensagens de uso

M E T H O D O

M E T H O D O

M E T H O D O

Zero

Test

M E T H O D O

M E T H O D O

M E T H O D O

M E T H O D O

M E T H O D O

M E T H O D O

0.0.0

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

Zero

Test

Mode

÷

Err

ou

 HI

EOI

-Err 

ou

 LO

LO BAT

ON

OFF

Cl

Cloro  0,05 - 6,0 mg/l

Ácido Cianúrico  2 - 160 mg/l

pH 6,5 - 8,4

0.0.0

Zero

Test

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

0.0.0

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

Cl

Cl

Zero

Test

Zero

Test

pH

P

Alcalinidade Total  0,1 - 4,0 mmol/l

0.0.0

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

Zero

Test

tA

Ligue o aparelho, pressionando a tecla ON/OFF.

No visor aparecerá o seguinte:

Seleccione o parâmetro que deseja medir, usando a tecla MODE:
Cl

 pH 

→ 

Cys

→ 

tA

→ 

CA.H

→ 

Cl

→ 

.......... (ciclo)

No visor aparecerá o seguinte:

Coloque uma célula limpa com água da amostra até à
marca de 10 ml, feche-a usando a tampa da célula e
coloque-a no compartimento de teste alinhando a marca

 da célula com marca 

Δ

 do aparelho.

Pressione a tecla ZERO/TEST.

O símbolo correspondente ao método piscará durante 3 segundos.

No visor aparecerá o seguinte:

Depois de efectuada a leitura do zero, retire a célula do
compartimento de teste.
A coloração aparecerá após a adição de um comprimido
do parâmetro que pretende medir.
Feche a célula novamente e coloque-a no compartimento
de teste com as marcas 

Δ

   alinhadas.

Pressione novamente a tecla ZERO/TEST.

O parâmetro a medir  piscará durante 3 segundos.

No visor aparecerá o seguinte:

Repetição de análises:

Pressione novamente a tecla ZERO/TEST.

Nova determinação do “zero”:

Pressione a tecla MODE até o parâmetro desejado
aparecer no visor.

Absorção de luz elevada. Motivo - exemplo: lente suja.

Intervalo de medição excedido ou excesso de turvação.

Resultado abaixo do intervalo de medição.

Substituir a pilha de 9 V. Não é possível fazer mais leituras.

Dureza Cálcica  50 - 500 mg/l CaCO

3

0.0.0

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

Zero

Test

CA.H

Utilize o adaptador para células de 16 mm.
Encha uma célula de 16 mm com 8 ml de água destilada
(sem Dureza). Adicione um comprimido CALCHECK e
esmague-o até se dissolver por completo usando um
triturador limpo. Feche a célula e coloque-a no adaptador
da célula (que por sua vez já está instalado no
compartimento de teste) com as marcas 

Δ

|

 alinhadas.

Pressione a tecla ZERO/TEST.

O símbolo tA piscará durante 3 segundos.

No visor aparecerá o seguinte:

Retire a célula do compartimento de testes. Adicione 2
ml da água da amostra. Feche a célula e coloque-a
novamente no compartimento de testes com as marcas
I e D  alinhadas.

Aguardar 2 minutos por uma reacção colorida.

Pressione a tecla ZERO/TEST.

O símbolo CA.H piscará durante 3 segundos.

O resultado aparecerá no visor, expresso em  mg/l CaCO

3

.

Tolerância:

±

 5% em toda a escala

Efectue a leitura do “Zero”. (Ver“manuseamento”).
Retire a célula do compartimento de teste, adicione um
comprimido ALKA-M-PHOTOMETER e esmague-o até
se dissolver por completo usando um triturador limpo.
Feche a célula e coloque-a no compartimento de teste
com as marcas  

Δ

 alinhadas.

Pressione a tecla ZERO/TEST.

O símbolo tA piscará durante 3 segundos.

O resultado aparecerá no visor, expresso em mg/l CaCO

3

.

Tolerância:

±

 5 % em toda a escala

Zero

Test

CA.H

•Cys

Zero

Test

•Cys

0.0.0

Zero

Test

•Cys

R E S U L T A D O

R E S U L T A D O

R E S U L T A D O

Adicione um comprimido CYANURIC ACID e esmague-
o usando um triturador limpo. O Ácido Cianúrico originará
um pouco de turvação com semelhança a leite. Feche a
célula e agite durante 20 até o comprimido se dissolver
completamente. Coloque a célula no compartimento de
teste com as marcas  

Δ

 alinhadas.

Pressione a tecla ZERO/TEST.

O símbolo Cys piscará durante 3 segundos.

O resultado aparecerá no visor, expresso em mg/l ácido
cianúrico.

Tolerância:

±

 10 mg/l

(a) Cloro livre

Efectue a leitura do “Zero”. (Ver “manuseamento”).
Esvazie a célula e em seguida adicione um comprimido
DPD No.1. Com o triturador limpo esmague o comprimido
e depois encha com a água da amostra até à marca dos
10 ml. Agite até o comprimido se dissolver completamente,
ponha a rolha na célula e coloque-a no compartimento
de teste com as marcas  

Δ

 alinhadas.

Pressione a tecla ZERO/TEST.

O símbolo “Cl”  piscará durante 3 segundos.

O resultado aparecerá no visor, expresso em mg/l de Cloro Livre.

(b) Cloro total

Retire a rolha e adicione um comprimido DPD No.3 à amostra
já previamente colorida após a medição anterior e  esmague-
a usando um triturador limpo. Agite até o comprimido se
dissolver completamente, feche a célula e coloque-a no
compartimento de teste com as marcas  

Δ

 alinhadas.

Aguardar 2 minutos por uma reacção colorida.

Pressione a tecla ZERO/TEST.

O símbolo “Cl”  piscará durante 3 segundos.

O resultado aparecerá no visor, expresso em em mg/l de Cloro Total.

Após cada medição, lave sempre a rolha.

(c) Cloro

Cloro combinado = Cloro total - Cloro livre

Tolerância:

0-1 mg/l: 

±

 0,05 mg/l > 3-4 mg/l: 

±

 0,30 mg/l

> 1-2 mg/l: 

±

 0,10 mg/l > 4-6 mg/l: 

±

 0,40 mg/l

> 2-3 mg/l: 

±

 0,20 mg/l

Retire a célula do compartimento de medição
e adicione um comprimido PHENOLRED /PHOTO-
METER à amostra. Esmague o comprimido até se
dissolver completamente, feche a célula e coloque-a no
compartimento de teste com as marcas  

Δ

 alinhadas.

Pressione a tecla ZERO/TEST.

O símbolo pH  piscará durante 3 segundos.

O resultado do pH aparecerá no visor.

Após cada medição, lave sempre a rolha.

Tolerância: 

±

 0,1 pH

No visor aparecerá o seguinte:

Encha numa célula limpa com 5 ml de água

da amostra. Volte a encher até a marca dos 10 ml com
água destilada.  Feche a célula e coloque-a no
compartimento de teste com as marcas  

Δ

 alinhadas.

Pressione a tecla ZERO/TEST.

O símbolo Cys piscará durante 3 segundos.

No visor aparecerá o seguinte:

Содержание 54568

Страница 1: ... the cap tightly and place the vial in the sample chamber making sure the Δ and marks are aligned Press the ZERO TEST key The method symbol flashes for approx 3 seconds The mg l CaCO3 value is shown in the display Rinse the vial and cap thoroughly after each test Tolerance 5 Full Scale Total Alkalinity 5 200 mg l CaCO3 0 0 0 R E S U L T R E S U L T Zero Test tA Switch the unit on using the ON OFF ...

Страница 2: ...ess both the MODE and ZERO TEST and keep them depressed Switch the unit on using the ON OFF key Release the MODE and ZERO TEST keys after approx 1 second The following messages will appear in turn on the display The calibration is reset to the factory setting SEL stands for Select or Calibration has been set by the user If the user calibration is to be retained switch the unit off using the ON OFF...

Страница 3: ...ondes Les résultat s affiche en mg l chlore total c Chlore combin Chlore combiné chlore total chlore libre Tol rance 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Procéder au calibrage du zéro voir Mise en service Ajouter une pastille de PHENOLRED PHOTOMETER dans l échantillon de 10 ml en la sortant directement de la feuille de protection sans la to...

Страница 4: ...de CAL Cl CAL RESULTAT x RESULTAT x Mode Zero Test 0 0 0 Zero Test R E S U L T A T R E S U L T A T CAL Zero Test On Off Remarque CAL cAL M E T H O D E M E T H O D E M E T H O D E M E T H O D E Valeures de calibrage recommand Chlore entre 0 5 et 1 5 mg l pH entre 7 6 et 8 0 Acide cyanurique entre 30 et 60 mg l Alcalinité totale entre 50 et 150 mg l CaCO3 Dureté calcique entre 100 et 200 mg l CaCO3 ...

Страница 5: ... Tolerancias 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Realizar la calibraci n a cero v ase instrucciones A los 10 ml de prueba acuosa a adir directamente de su envoltura una tableta PHENOLRED PHOTOMETER y machacarla con una varilla limpia Disolver completamente la tableta cerrar la cubeta con su tapa y posicionarla Δ Presionar ZERO TEST El s mb...

Страница 6: ...roduce errores de medici n En este caso se tapa la cubeta y las burbujas se disuelven girando la cubeta antes de realizar el an lisis 7 La infiltraci n de agua en la c mara de medici n debe ser evitada La entrada de agua en la caja del fot metro puede destruir las piezas de construcci n electr nicas y producir da os de corrosion 8 El ensuciamiento de la ptica diodo luminoso y fotosensor en la c ma...

Страница 7: ... de método presionar la tecla MODE CAL Cl CAL pH CAL Cys CAL S 4 3 Scroll Realizar la calibración a cero como descrito Presionar la tecla ZERO TEST El símbolo del método parpadea durante aproximadamente 3 segundos En el display aparece alternativamente Colocar el estándar a utilizar en el compartimento de medición según las marcas Presionar la tecla ZERO TEST El símbolo del método parpadea durante...

Страница 8: ...y appare il risultato in mg l del cloro totale c Cloro combinato Cloro legato cloro totale cloro libero Tolleranza 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Condurre la taratura a zero vedere la messa in funzione Aggiungere direttamente nei 10 ml dell acqua di prova una compressa di PHENOLRED PHOTOMETER senza toccarla con le dita e frantumarla c...

Страница 9: ... e le bollicine si dissolvono agitando la cuvetta prima di procedere al test 7 Si deve impedire la penetrazione di acqua nel pozzetto di misurazione L entrata dell acqua nel corpo del fotometro pu comportare un danneggiamento dell elettronica e danni dovuti a corrosione 8 L apparecchiatura ottica diodo luminoso e fotosensore nel pozzetto di misurazione se sporca comporta errori di misurazione Le s...

Страница 10: ...ffettuata dall operatore scorretto cancellato E 10 E 70 E 72 E 74 E 71 E 73 E 75 E 76 E 77 E 79 E 78 Calibrazione effettuata dall operatore cAL Calibrazione effettuata in fabbrica CAL Zero Test Mode On Off CAL SEL SEL CAL Mode On Off cAL SEL Indicazione Metodo di calibrazione On Off Mode CAL Cl CAL RISULTATO x RISULTATO x Mode Zero Test 0 0 0 Zero Test R I S U L T A T O R I S U L T A T O CAL Zero ...

Страница 11: ...rdar 2 minutos por uma reacção colorida Pressione a tecla ZERO TEST O símbolo CA H piscará durante 3 segundos O resultado aparecerá no visor expresso em mg l CaCO3 Tolerância 5 em toda a escala Efectue a leitura do Zero Ver manuseamento Retire a célula do compartimento de teste adicione um comprimido ALKA M PHOTOMETER e esmague o até se dissolver por completo usando um triturador limpo Feche a cél...

Страница 12: ...itura Se houver suspeita de contaminação examine o visor 9 Quando usar reagentes em comprimidos use apenas comprimidos protegidos com folha metálica No caso dos comprimidos PHENOLRED devem estar marcados na folha metálica com PHOTOMETER 10 Os comprimidos devem ser adicionados à amostra de água sem serem tocados com os dedos 11 Elevadas diferenças de temperatura entre o fotómetro e o ambiente de tr...

Страница 13: ...n de display verschijnt het meetresultaat in mg l totaal chloor c gebonden Chloor gebonden chloor totaal chloor vrij chloor Toleranties 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Doe de nulstelling zie In bedrijf nemen Doe een PHENOLRED PHOTOMETER tablet direkt uit de folie in 10 ml monsterwater en maak deze fijn met een schone roerstaaf Tablet g...

Страница 14: ...igingen voorbehouden Printed in Germany 09 00 Druk op de MODE toets en houd deze ingedrukt Schakel de fotometer in met de ON OFF toets laat na ca 1 seconde de MODE toets los Druk voor het wisselen van het meetbereik de MODE toets CALCl CALpH CALCys CAL tA CALCA H Scroll Doe de normale nulstelling Druk de ZERO TEST toets Het methode symbool knippert ca 3 seconden In de aanwijzing verschijnt afwisse...

Страница 15: ...EST Metodesymbolet blinker i ca 3 sekunder Displayet viser resultatet i mg l klor total c bundet klor bundet klor klor total frit klor Tolerance 0 1 mg l 0 05 mg l 3 4 mg l 0 30 mg l 1 2 mg l 0 10 mg l 4 6 mg l 0 40 mg l 2 3 mg l 0 20 mg l Gennemfør nulkompensation se ibrugtagning En PHE NOLRED PHOTOMETER tablet tilsættes til 10 ml vandprøven direkte fra folien og knuses med en ren rørepind Opløs ...

Страница 16: ...tiveres fabrikations kalibreringen Displayet viser skiftevis Apparatet slukkes på tasten ON OFF Kalibrering out of range Cl Fabrikations kalibrering ukorrekt slettet pH Fabrikations kalibrering ukorrekt slettet Cys Fabrikations kalibrering ukorrekt slettet TA Fabrikations kalibrering ukorrekt slettet CAH Fabrikations kalibrering ukorrekt slettet Cl Bruger kalibrering ukorrekt slettet pH Bruger kal...

Отзывы: