SECTION 10 APPENDIX I
_______________________________________________________________________
10-5
10.4
Safety Tester Connection with PIC
This procedure describes a method to connect an BIS VISTA Monitoring System to a safety
tester. It uses a current date sensor to provide a contact point for the safety tester leads that
correspond to the patient contact points of the BIS VISTA system.
1.
Connect BISx (Digital Signal Converter) cable to BIS VISTA monitor.
2.
Connect PIC (Patient Interface Cable - sensor) cable to other end of BISx.
3.
Verify sensor for test is of current date code. Connect sensor to PIC cable.
4.
Remove gel and green pad from each sensor contact point.
5.
Connect safety tester leads per tester instructions to sensor contact point pads (i.e.
via alligator clips).
6.
Apply power to the BIS VISTA system per safety tester instructions. Make required
leakage measurements.
7.
Disconnect and dispose of sensor. NOTE: DO NOT discard PIC or BISx. These
are re-usable components.
Figure 13 - Safety Tester Contact Points
BIS
VISTA
M i
BISx
Patient Interface
Cable (PIC)
Contact points for
safety tester.
See steps 4 and 5.