SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-8
Status Messages: Causes:
Corrective Actions:
96
Sensor Invalid
1.
Poor or contaminated
connection between
sensor and PIC.
2.
Defective sensor.
3.
Defective PIC.
4.
Defective BISx.
1.
Connect/clean connection
between sensor and PIC.
2.
Replace the sensor.
3.
Replace the PIC.
4.
Replace BISx Bulkhead
Connector or replace BISx.
109
Sensor Used for Over
24 Hours
Sensor has been sending
data for more than 24 hours
Replace sensor
202
Sensor Expired
1.
Sensor expiration date
has passed
2.
Time and date are
incorrect
1.
Replace Sensor
2.
Verify current time and
date; Reset if necessary
1000-1999
Unrecoverable BISx
Error
1.
Poor connection
between BISx monitor
cable and monitor.
2.
Defective BISx
3.
Defective monitor.
1.
Unplug BISx from monitor
and plug in again.
2.
Replace the BISx or the
Monitor Interface Cable.
3.
Replace the monitor
2000-2999
Unrecoverable
Monitor Error
A system error has
occurred. The monitor may
stop operating.
1.
Turn the monitor off, then
on again.
2.
Reset monitor.
3.
Replace monitor.
3000-3999
Data Export Error
The data export was not
successful.
1.
Removable drive is not
connected properly to
USB(A) port.
2.
Drive is incompatible or
defective.
3.
Drive is full.
1.
Check connection.
2.
Replace drive.
3.
Replace drive.
4000-4999
Software Update Error
A software error has
occurred.
1.
Restart monitor.
2.
Reset monitor.
3.
Replace monitor.
5000-5999
Trend Memory Error
A memory error has
occurred.
1.
Restart monitor.
2.
Reset monitor.
3.
Replace monitor.