SECTION 9 SPECIFICATIONS AND WARRANTY
_______________________________________________________________________
9-2
EEG Specifications:
Epoch Duration:
2 seconds
Artifact Rejection:
Automatic
EEG Scales:
25
µ
V/div (+/- 1 mV Full Scale)
EEG Sweep Speeds:
25 mm/sec
Computed Parameters:
Bispectral Index, Suppression Ratio, EMG, Signal
Quality Index, and Burst Count
User-defined Displays:
Trend and real-time EEG waveforms
Update Rate:
1 second for BIS number, 10 seconds for Trend
Event Markers:
User selected
Alarms:
Auditory and visual, user adjustable limits
Filters:
ON (2 – 70 Hz with notch) or OFF (.25 – 100 Hz)
Mode:
Sensor automatically selects mode.