SECTION 1 SAFETY PRECAUTIONS
________________________________________________________________________
1-2
FOR PROPER GROUNDING, THE POWER RECEPTACLE MUST BE A
THREE-WIRE GROUNDED OUTLET. A HOSPITAL GRADE OUTLET IS
REQUIRED. NEVER ADAPT THE THREE-PRONG PLUG FROM THE
MONITOR TO FIT A TWO-SLOT OUTLET. IF THE OUTLET HAS ONLY
TWO SLOTS, MAKE SURE THAT IT IS REPLACED WITH A THREE-SLOT
GROUNDED OUTLET BEFORE ATTEMPTING TO OPERATE THE
MONITOR.
IF THE INTEGRITY OF THE EXTERNAL PROTECTIVE EARTH GROUND
IS IN DOUBT, THE BIS VISTA SYSTEM SHALL BE OPERATED FROM ITS
INTERNAL BATTERY POWER SOURCE ONLY.
FOR BIS VISTA SYSTEMS USED OUTSIDE OF NORTH AMERICA: A
HARMONIZED LINE CORD WITH CONDUCTORS HAVING A CROSS
SECTIONAL AREA GREATER THAN 0.75 mm
2
MUST BE USED.
BE SURE THE MONITOR IS MOUNTED SECURELY IN PLACE TO AVOID
PERSONAL OR PATIENT INJURY.
UNIVERSAL PRECAUTIONS SHALL BE OBSERVED TO PREVENT
CONTACT WITH BLOOD OR OTHER POTENTIALLY INFECTIOUS
MATERIALS. PLACE CONTAMINATED MATERIALS IN REGULATED
WASTE CONTAINER.
WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS
OCCURS, RE-TEST BEFORE FURTHER USE.
DO NOT MIX DISINFECTING SOLUTIONS (e.g., BLEACH AND AMMONIA),
AS HAZARDOUS GASES MAY RESULT.
ELECTRICAL SHOCK HAZARD: THE MANUFACTURER'S INSPECTION OF
THIS APPARATUS VERIFIED THAT THE GROUND LEAKAGE CURRENT
AND THE PATIENT SAFETY CURRENT WERE LESS THAN THE
SPECIFIED LIMITS ESTABLISHED BY THE APPLICABLE SAFETY
STANDARDS. AS A MATTER OF SAFE PRACTICE, THE INSTITUTION
SHOULD CONDUCT PERIODIC TESTS TO VERIFY THESE CURRENTS.
ELECTRICAL SHOCK HAZARD: DO NOT REMOVE MONITOR COVERS
DURING OPERATION OR WHILE POWER IS CONNECTED TO MONITOR.
GROUND WIRE LEAKAGE CURRENT MUST BE CHECKED WHENEVER
INSTRUMENT CASE IS OPENED BY A QUALIFIED BIOMEDICAL
ENGINEERING TECHNICIAN.
POWER SUPPLY IS INTERNALLY FUSED. REPLACE POWER SUPPLY ONLY
WITH ASPECT MEDICAL SYSTEMS BIS VISTA POWER SUPPLY.
Содержание BIS VISTA
Страница 2: ......
Страница 4: ......
Страница 32: ...SECTION 3 PRINCIPLES OF OPERATION _______________________________________________________________________ 3 10...
Страница 48: ...SECTION 6 PREVENTIVE MAINTENANCE _______________________________________________________________________ 6 8...
Страница 90: ......
Страница 93: ......