INTRODUCTION
This section provides instructions for replacing the PIC, BISx, and the monitor, and for
removing and replacing parts of the monitor and the BISx. If a component needs to be
serviced, please consult Sections 8.12 and 8.13 for instructions on packaging and shipping.
The BIS VISTA Monitoring System is designed to be easily serviced by using the built in
diagnostic routines (see Section 7 “Diagnostics and Troubleshooting”) and the major
component swapping techniques described below. Replacement part numbers are listed in
Section 10.1, “Accessories and Spare Parts List.”
8.1
Replacing the PIC
Parts Required:
Tools Required:
Patient Interface Cable (PIC Plus) P/N 186-0107
None
To replace the PIC, unplug the PIC cable from the BISx Bulkhead Connector by grasping
the connectors (NOT the cable!) and firmly pulling the two sections apart.
To attach the replacement PIC cable to the BISx, align and press the PIC and BISx
connectors together firmly.
8.2
Replacing the BISx
Parts Required:
Tools Required:
BIS VISTA BISx with Patient Interface Cable;
P/N 185-0145-AMS
None
To replace the BISx:
Disconnect the BISx from the front of the monitor. To do this, carefully grasp the
connector on the BISx Monitor Interface Cable and pull. DO NOT twist or pull on the
cable.
If necessary, unplug the PIC cable from the BISx by grasping the connector (NOT the
wires!) and firmly pulling it out of the BISx Bulkhead Connector.
To install the replacement BISx, hold the cylindrical connector with the flat side up (at 12
o’clock position) and insert firmly into BISx port on front of monitor.
To re-attach the PIC cable, align the PIC and BISx connectors and press together firmly.