Guidelines and manufacturer's declaration
– Electromagnetic immunity (Table 204)
The product is intended for operation in the electromagnetic environment specified below. The customer / user
of the product must ensure that it is used in such an environment.
Immunity tests
IEC 60601-
checking rule
Compliance Level
Electromagnetic Environmental Guidelines
Realized HF IEC 61000-
4-6
3 Vrms
150 kHz bis
80 MHz
10 Vrms
Portable and mobile RF
Communication Devices
should not be used at a distance from the
product, including the cables, than the
recommended protective distance according
to the applicable for the transmission
frequency
Equation is calculated.
Radiation-HF IEC
61000-4-3
3 V/m
80 MHz bis
2,5 GHz
3 V/m
Recommended protective distance
d=0,35√P
d=1,2√P 80 MHz to 800 MHz
d=2,3√P 800 MHz to 2,5 GHz
With P than the rated power of the
transmitter in watts (W) according to the
transmitter manufacturer and d as a
recommended protective distance in meters
(m).
The field strength of stationary radio
transmitters determined by electromagnetic
location measurement, a
–
should be lower than the CONFORMITY
LEVEL in all frequency ranges
be. b In the environment of devices that
have the following symbol, disturbances are
possible:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: This guide may not be applicable in all situations. The propagation of electromagnetic waves is
determined by absorption and reflections by structures,
Objects and people influenced.
a The field strength of stationary transmitters, such as base stations of radio telephones and mobile land radios,
amateur radio stations, AM and FM transmitters as well as radio and television stations cannot theoretically be
precisely predetermined. In order to determine the electromagnetic environment due to the fixed RF transmitters,
site-side electromagnetic monitoring is recommended. If the measured field strength in the environment in which
the product is to be used exceeds the applicable RF compliance level, special care must be taken to ensure that
normal operation of the product can be ensured. If abnormalities are detected, additional measures may be
required, such as a different orientation or a change of location of the product.
b In the frequency range from 150 kHz to 80 MHz, the field strength should be less than 10 V/m.
Содержание motioncare LIANA
Страница 15: ...Gebrauchsanweisung LIANA Version 1 0 Seite 15 von 50 7 1 Product dimensions...
Страница 39: ...Gebrauchsanweisung LIANA Version 1 0 Seite 39 von 50 21 ELECTRICAL DIAGRAM...
Страница 40: ...Gebrauchsanweisung LIANA Version 1 0 Seite 40 von 50...
Страница 42: ...Gebrauchsanweisung LIANA Version 1 0 Seite 42 von 50 23 CE DECLARATION OF CONFORMITY PSS LIANA...
Страница 50: ...Gebrauchsanweisung LIANA Version 1 0 Seite 50 von 50...