Guidelines and manufacturer's declaration
– Electromagnetic immunity (Table 202)
The product is intended for operation in the ELECTROMAGNETIC ENVIRONMENT specified below. The
customer or the user of the product should ensure that it is used in such an environment.
Immunity tests
IEC 60601- checking rule
Compliance Level
Electromagnetic
Environmental Guidelines
Discharge of static
electricity (ESD)
according to IEC 61000-
4-2
± 6kV contact discharge
± 8kV air discharge
± 6kV contact
discharge
± 8kV air discharge
The floor should be made
of wood, concrete or
ceramic tiles. If the floor is
provided with synthetic
material, the relevant
humidity must be at least
30% amount to.
Fast transient
interference pulses/Burst
IEC 61000-4-4
± 2 kV for power supply
line
± 1 kV for Input/output lines
± 2 kV for power
supply line
Omitted for
input/output lines
The quality of the supply
voltage should be that of a
typical business or
Match hospital
environment.
Overvoltage IEC 61000-
4-5
±/1 kV Line against line
±/2 kV Line against
grounding
±/1 kV Line against
line
±/2 kV Line against
grounding
The quality of the supply
voltage should match that
of a typical business or
hospital environment.
Voltage drops, short
interruptions
and voltage fluctuations
for the power supply
input lines IEC 61000-4-
11
<5 % UT
(>95 % Collapse of the UT)
for 0,5 Period
40 % UT
(60 % Collapse of the UT)
for 5 Period
70 % UT
(30 % Collapse of the UT)
for 25 Period
<5 % UT
(>95 % Collapse of the UT)
for 5 s
<5 % UT
(>95 % Collapse of
the UT)
for 0,5 Period
40 % UT
(60 % Collapse of the
UT) for
5 Period
70 % UT
(30 % Collapse of the
UT) for
25 Period
<5 % UT
(>95 % Collapse of
the UT)
for 5s
The quality of the
Supply voltage
should the one
Correspond to typical
business or hospital
environment.
Current frequency (50/60
Hz)
Magnetic field IEC
61000-4-8
3 A/m
3 A/m
Magnetic fields at the
mains frequency should
match the typical values
found in the business or
hospital environment,
conform.
ATTENTION UT is the ac mains voltage before the application of the test stage.
Содержание motioncare LIANA
Страница 15: ...Gebrauchsanweisung LIANA Version 1 0 Seite 15 von 50 7 1 Product dimensions...
Страница 39: ...Gebrauchsanweisung LIANA Version 1 0 Seite 39 von 50 21 ELECTRICAL DIAGRAM...
Страница 40: ...Gebrauchsanweisung LIANA Version 1 0 Seite 40 von 50...
Страница 42: ...Gebrauchsanweisung LIANA Version 1 0 Seite 42 von 50 23 CE DECLARATION OF CONFORMITY PSS LIANA...
Страница 50: ...Gebrauchsanweisung LIANA Version 1 0 Seite 50 von 50...