8.
System Setup Menu:
Press and hold the "CAL" button for 1 second (until you hear a double beep) to enter the System
Setup menu.
After entering the System Setup Menu, the following setup options will be displayed on the LCD
Display.
Do you want to recall factory settings?
Press: 1=No, 2=Yes, Mode=Escape
Recalling the factory settings allows you to reset your Aseptico VCT to factory specifications. Please
note, any customized presets will be lost if you recall factory settings.
To keep your customized settings, press 1.
To return the Aseptico VCT’s preset buttons to the factory settings, press 2.
To exit the system setup menu without saving changes, press Mode.
Do you want to upload new software?
Press: 1=No, 2=Yes, Mode=Escape
New versions of software can be downloaded from the internet with a PC then uploaded onto the
Aseptico VCT using the included serial port connector cable. Use your Personal Computer to down-
load the AseptiCom serial port communications program from www.aseptico.com before beginning.
To keep the current software version, press 1.
To upload new software, connect the serial cable from a PC to the connector labeled "Serial" on the
back of the Aseptico VCT, then press 2. The display will read:
Ready for upload. Settings = 57600,8,N,1
Press Mode to escape
Do you want the warning beeper enabled?
Press: 1=Yes, 2=No, Mode=Escape
When the warning beeper is active, a chirping warning noise will be audible when the torque level
reaches approximately 75% of the selected torque limit.
To enable the warning beeper, press 1.
To disable the warning beeper, press 2.
To exit the system setup menu without changes, press Mode.
Do you want the reverse beeper enabled?
Press: 1=Yes, 2=No, Mode=Escape
When the reverse beeper is active, an audible beep will sound when the handpiece is rotating in the
reverse direction.
To enable the Reverse Beeper feature, press 1.
To disable the Reverse Beeper feature, press 2.