Aseptico penguin rfa Скачать руководство пользователя страница 4

 

4

 

 
 
 
 
 
 
 
 
 
9. Technical function 
For bringing the measurement pin into vibration, short magnetic pulses are sent from the instrument tip. The magnetic pulses 
interact with the magnet inside the measurement pin and cause the pin to vibrate. A pickup in the instrument picks up the alternating 
magnetic field from the vibrating magnet, calculate the frequency and from that, the ISQ value. 
 
10. Implant stability 
An implant can have different stabilities in different directions. Make sure to measure from different directions around the top of the 
MulTipeg. 
 
11. ISQ-value 
The stability of the implant is presented as an “ISQ value”. The higher the value, the more stable the implant. The ISQ is described in 
numerous clinical studies. A list of studies can be ordered from the supplier. 
 
12. Batteries & charging 
The instrument contains 2 NiMh battery cells that have to be charged before use. A full charge take approximately 3 hours. From fully 
charged, the instrument can measure continuously for 60 minutes before it needs to be recharged. The yellow LED is lit when the 
battery needs recharging. When the battery reaches a critical level, the instrument shuts off automatically. When the batteries are 
charging, the blue LED is lit. The charger should not be plugged in while measuring due to the risk of power line interference making it 
difficult to measure. 
 
13. Usage 
 
13.1 Instrument on/off 
To turn the instrument on, press the operating key. A short beep should be heard and then all display segments are lit up for a short 
while. Check that all display segments are lit.  
The software version is then shown briefly before the instrument starts to measure. If any error code (EX, where “X” is the error 
number) is shown during start up, please refer to the section “Troubleshooting”. 
To turn off, press and hold the operating key until the instrument turns off. The instrument will power down automatically after 1 
minute of inactivity. 
 
13.2 Measurement 
A MulTipeg is mounted onto the implant by using the MulTipeg driver. Use hand-tightening (6-8 Ncm) of tightening force. Turn on the 
instrument and hold the tip close to the top of the pin (fig 5). When a signal is received, a buzzing sound is heard and then the ISQ-
value is shown on the display for a short while before the instrument starts to measure again. 
If electromagnetic noise is present, the instrument cannot measure. The electromagnetic noise warning is audible as well as visible on 
the display. Try to remove the source of the noise. The source could be any electric equipment close to the instrument.  
 
14. Cleaning and maintenance 
 
14.1 Recommended disinfectants 
The following disinfectants are recommended for the instrument, the MulTipeg and the MulTipeg Driver: 

• 

Schülke & Mayr: Mikrozid AF Liquid 

• 

Dürr: AF 322 

• 

Metrex: CaviCide 

 
Follow the instructions for use for the disinfectant that is used. 
 
14.2 Autoclave sterilization (MulTipeg & MulTipeg Driver) 
Sterilization should be made in a pre-vacuum steam sterilizer (autoclave) according to ISO 17665-1. 
Clean the products and put them in an FDA-cleared autoclave bag before sterilization. 
The following sterilization process shall be used: 
 

 

The instrument emits short magnetic pulses with pulse duration of 1 ms and strength of +/- 20 gauss, 10 mm 
from the instrument tip. Precautions might be necessary when using the instrument close to cardiac 
pacemakers or other equipment sensitive to magnetic fields. 
 

 

Before first use, the parts should be cleaned and disinfected. 

 

Содержание penguin rfa

Страница 1: ...1 PenguinRFA English 3 Deutsch 6 Espa ol 9 Fran ais 12 Italiano 15 Portugu s 18 Svenska 21 Integration Diagnostics Sweden AB Nederg rdsgatan 5 416 54 G teborg Sweden...

Страница 2: ...2 Fig 1 Fig 2 Fig 3 Fig 4 Fig 5...

Страница 3: ...deral law restricts this device to sale by or on the order of a physician or dentist 7 Characteristics Penguin RFA is an instrument for measuring the stability ISQ of dental and craniofacial implants...

Страница 4: ...starts to measure If any error code EX where X is the error number is shown during start up please refer to the section Troubleshooting To turn off press and hold the operating key until the instrume...

Страница 5: ...t to bring the MulTipeg into vibration If so try to hold the instrument tip closer to the tip of the peg Check also that no soft tissue is touching the peg which might stop its vibration Noise warning...

Страница 6: ...inem Zahnarzt oder im Auftrag eines Zahnarztes oder Arztes gekauft werden 7 Merkmale PenguinRFA ist ein Ger t zur Messung der Stabilit t ISQ von Zahn und kraniofazialen Implantaten es misst die Resona...

Страница 7: ...oftwareversion kurz angezeigt bevor das Instrument zu messen beginnt Wenn beim Hochfahren Fehlercodes angezeigt werden EX wobei X die Fehlernummer angibt lesen Sie bitte im Abschnitt Fehlersuche nach...

Страница 8: ...iger f r den PenguinRFA sein den MulTipeg in Schwingung zu versetzen Sollte dies der Fall sein bringen Sie die Spitze des PenguinRFA n her an die Spitze des MulTipeg heran Pr fen Sie au erdem dass die...

Страница 9: ...estringen la venta de este dispositivo a trav s bajo las rdenes de un m dico o dentista 7 Caracter sticas PenguinRFA es un instrumento para medir la estabilidad ISQ de los implantes dentales y craneof...

Страница 10: ...ados Posteriormente se muestra brevemente la versi n de software antes de que el instrumento comience la medici n Si aparece un c digo de error EX donde X es el n mero de error durante el encendido co...

Страница 11: ...hacer vibrar al MulTipeg Si este fuera el caso trate de mantener la punta del instrumento m s cerca de la punta del MulTipeg Adem s compruebe que ning n tejido blando est tocando el MulTipeg ya que es...

Страница 12: ...rale limite cet appareil la vente par ou sur ordonnance d un m decin ou d un dentiste 7 Caract ristiques Penguin RFA est un instrument de mesure de la stabilit ISQ des implants dentaires et craniofaci...

Страница 13: ...mence la mesure Si un code d erreur EX o X est un num ro d erreur s affiche pendant le d marrage veuillez vous r f rer la section D pannage Pour l arr ter appuyer et maintenir enfonc e la touche de fo...

Страница 14: ...nt d amener le MulTipeg en vibration Le cas ch ant essayer de tenir la pointe de l instrument pr s de la pointe du peg V rifier galement qu aucun tissu mou ne touche le peg ce qui pourrait arr ter sa...

Страница 15: ...ntisti 8 Caratteristiche PenguinRFA uno strumento per la misurazione della stabilit ISQ di impianti dentali e craniofacciali Lo strumento misura la frequenza di risonanza di un pin di misurazione dett...

Страница 16: ...gmenti del display si illumineranno brevemente Viene brevemente visualizzata la versione del software prima che lo strumento inizi la misurazione Se viene visualizzato un codice errore EX dove X il nu...

Страница 17: ...li in autoclave prima di essere degradato 16 Risoluzione dei problemi 16 1 Possibili errori Difficolt ad ottenere una misurazione In alcuni casi pi difficile per lo strumento portare il MulTipeg in vi...

Страница 18: ...Rx Only A venda deste dispositivo est restringida a dentistas ou por ordem de um m dico ou dentista 7 Caracter sticas Penguin RFA um instrumento para a medi o da estabilidade ISQ de implantes dent ri...

Страница 19: ...se de que todos os segmentos do visor acendem A vers o do software apresentada brevemente antes de o instrumento iniciar a medi o Se for apresentado algum c digo de erro EX sendo X n mero de erro dura...

Страница 20: ...veis erros Dificuldade em obter uma medi o Em alguns casos mais dif cil para o instrumento colocar o MulTipeg em vibra o Se este for o caso tente segurar a ponta do instrumento mais pr ximo da ponta...

Страница 21: ...v ndas av kliniskt utbildad personal 7 Beskrivning PenguinRFA r ett instrument avsett att m ta stabiliteten p dentala och craniofaciala implantat Instrumentet m ter resonansfrekvensen hos en m tpinne...

Страница 22: ...n instrumentet b rja m ta Om ett felmeddelande visas EX d r X r felnumret se sektionen Fels kning F r att st nga av tryck ner knappen igen Instrumentet st ngs automatiskt av efter 1 minuts inaktivitet...

Страница 23: ...peg att vibrera Om det r s f rs k att h lla nosen p instrumentet n rmare toppen p MulTipeg Kontrollera ocks att ingen mjukv vnad r r MulTipeg vilket kan d mpa dess vibration Brusvarning ljudsignal och...

Страница 24: ...24 Integration Diagnostics Sweden AB Nederg rdsgatan 5 416 54 G teborg Sweden 15013 03 Ifu ID...

Отзывы: