25
The AEU-27A has the ability to load
software updates and enhance the
functionality of the system, should this be
desired. A card slot, labeled “Memory Card
Port”, is provided on the back of the unit
(see Figure 7). This Port accepts memory
cards, P/N: MC-27A, that are very similar
to those used in common consumer
devices. These cards, available from
Aseptico, enable a user to update software
or replace existing software that might
have been accidentally erased or
corrupted. Contact Aseptico for more
information on card usage and availability.
To reprogram an AEU-27A unit, follow the
Steps below:
Programming Steps:
1
. Turn ‘Off’ the Main Power Switch on the
back panel.
2.
Grasp the right-hand end of the rubber
dust cover for the Memory Card Port and
pry open the cover to expose the card slot.
3.
Insert the new memory card in the slot with
label facing upward (card terminals should
face downward). Carefully and slowly press
card inward until a ‘click’ is felt, then
release card.
4.
Turn the Main Power Switch (on the back
panel) ‘On’.
5.
The AEU-27A Display will show the
following message:
• Press the RATIO key on the Control Panel.
6
. The Display will show the following message:
• Press the RATIO key on the Control Panel.
7.
The Display will show the following message:
• A status bar will indicate the progress of
the programming.
8
. When the programming is complete, the
Display will show the following message:
• Press the card inward slightly, then
release it to eject it. When the card is
ejected, the AEU-27A System will reset
with normal power-up screen displayed.
9.
Remove the memory card and store it in a
safe place. Close the rubber dust cover on
the Memory Card Port.
In the event that the programming
procedure is interrupted, the unit will
display the following message:
Then:
Re-start the programming procedure from
Step #1 (Remember to turn main power
‘Off’ before reprogramming).
REPROGRAMMING THE UNIT
Fig. 7 - Programming Slot
Memory Card Detected.
Re-program?
(RATIO = YES, CAL = NO)
Series Will Be Erased!
Continue?
(RATIO = YES, CAL = NO)
Programming...
Programming Successful.
Eject Card.
Programming Failed.
Eject Card.
Console Software Error.
Re-program unit.
MEMORY
CARD
DUST
COVER
Содержание AEU-27A