prompt will display:
The unit will run for approximately 30
seconds, measuring the handpiece for an
accurate display of speed and torque
values. It is recommended to calibrate at
the beginning of each day and when
handpieces are changed.
If calibration is successful, the following
prompt will appear, then the unit will return
to the last settings used:
If calibration fails, the following prompt will
appear advising the user to check the
motor/handpiece for problems. Press the
CAL button to escape out of the test:
B. System Setup:
Pressing, holding, then releasing the
CAL/SETUP button will enable the
System’s Setup options menu. Refer to
the System Setup section, page 16 for
complete instructions.
4 SPEED Buttons
Allow user to select desired
speed (RPM) for
motor/handpiece.
Press the SPEED Selector
Up/Down buttons to adjust
speed. The unit will
automatically enter MANUAL
mode and display the new
RPM settings with the ratio and torque
settings (example only):
Note:
For speed accuracy, the Ratio
Selector must match the ratio of the
handpiece being used.
Note:
Speed ranges are accurate when
the handpieces referenced on page 1 are
used with the supplied 30K motor. Speed
ranges are shown in Chart 2, page 10.
5 Forward/Reverse (FWD/REV)
Button:
Sets the rotational direction of
the motor/handpiece.
Press and release FWD/REV
button until the desired rotation is
indicated: the green LED illuminates when
forward rotation is selected; the amber
LED indicates reverse rotation. An audible
tone will also indicate when reverse
rotation is activated. This reverse audible
tone may be adjusted or turned off as
described in the Setup Options Section,
page 17.
6 Torque Adjustment Buttons:
Allow the user to adjust torque limits in
Gram•Centimeter increments.
Press the TORQUE Selector
Up/Down buttons to adjust
torque. The unit will
automatically enter MANUAL
MODE and display the new
TORQUE settings with the
ratio and speed settings
(example only):
11
MANUAL MODE
16:1 1800 RPM 60 g-cm
Calibration Failed!
Check Motor and Retest.
Press “CAL” to Escape.
Calibration Successful!
Please Wait.
MANUAL MODE
16:1 1800 RPM 90 g-cm
Testing In Progress...
Please Wait.