17

If pressure is then reapplied to the foot

pedal, the handpiece will return to forward

rotation.
When the Manual Reverse mode is

selected, the menu automatically returns to

the “Setup Options” menu and displays

“(MAN)” at the “Auto Stop Rev.” prompt as

shown below:

• Auto Stop Off

- When Auto Stop Off is

enabled and the foot pedal is activated, the

handpiece will continually rotate in the

forward direction and apply the selected

torque limit until the foot pedal is released.

• Cancel

- Exits out of the “Auto Stop

Reverse” menu and returns to the previous

Setup Options menu without saving any

changes.

C. Reverse Beeper:

When the Reverse

Beeper feature is enabled, an audible beep

will sound while the handpiece is rotating in

the reverse direction. After pressing the

SELECT button, the following menu will

display:

• On

- Enables the Reverse Beeper sound.

“Reverse Beeper (ON)” will be displayed at

the Setup Options menu. 

NOTE:

“Reverse

Beeper (ON)” is the factory default.

• Off

- Disables the Reverse Beeper

sound. No sound will be emitted while the

handpiece rotates in reverse. “Reverse

Beeper (OFF)” will be displayed on the

Setup Options menu. 

• Cancel 

- Exits out of the menu and

returns to the Setup Options menu without

saving any setup changes

D. Torque Warning:

When the Torque

Warning feature is active (“ON)”, a chirping

sound will be audible when the torque level

reaches approximately 75% of the selected

torque limit. When this option is selected,

the following menu will display:

• On

- Enables the Torque Warning sound.

“Torque Warning (ON)” will be displayed on

the Setup Options menu. 

Note:

“Torque

Warning (ON)” is the factory default.

• Off

- Disables the Torque Warning sound.

No tone will be emitted when the torque

reaches 75%. “Torque Warning (OFF)” will

be displayed on the Setup Options menu. 

• Cancel

- Exits out of the menu and

returns to the Setup Options menu without

saving any setup changes.

E. Accessory Warning Tone:

When the

Accessory Warning Tone feature is active

(“ON)”, a warning tone will be audible when

a GT

®

Accessory file, ITR Emulation

Accessory file, or ProTaper

®

Retreatment

file is in use. When this option is selected,

the following menu will display:

• On 

- Enables the Accessory Warning

Tone. A warning sound will be emitted

when the above accessory files are used.

“Accessory Warn (ON)” will be displayed on

the Setup Options menu.

Setup Options:

> Auto Stop Rev.   (MAN)

(etc...)

Reverse Beeper:

> On

Off

Cancel

Torque Warning:

> On

Off

Cancel

Accessory Warning Tone:

> On

Off

Cancel

Содержание AEU-27A

Страница 1: ...AEU 27A Electronic Endodontic System OPERATION AND MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...E OF CONTENTS Introduction 1 Package Contents 1 Indications for Use 2 Contraindications 2 Adverse Reactions 2 Warnings Safety Precautions 2 Step By Step Instructions 4 Setting Up the Unit 4 Quick Star...

Страница 3: ...may be provided to cover the operation and maintenance of handpieces or other accessories for your unit PACKAGE CONTENTS AEU 27A Electronic Control Console Autoclavable Brushless Micromotor P N AE 25...

Страница 4: ...rty manufacturers is the sole responsibility of the user CAUTION All repairs are to be performed by authorized Aseptico service personnel only WARNING Always follow these guidelines when operating the...

Страница 5: ...only WARNING Do not disassemble or alter the AE 250 30 micromotor console or foot switch CAUTION Use only appliance cord Type C13 10A per IEC EN 60320 1 Note North America Denmark Australia and New Ze...

Страница 6: ...Figure 3 The cord connector and receptacle are keyed to engage one another Insert then turn the locking nut clockwise to secure 4 Attach the remote power cord to the back of the console see Figure 3 a...

Страница 7: ...appropriate E Type handpiece to the motor as shown in Figure 4 Fig 3 Console Back POWER CORD IN MAIN POWER ON OFF SWITCH MEMORY CARD PORT FOOT SWITCH IN FUSES E TYPE HANDPIECE MOTOR Fig 4 Handpiece Mo...

Страница 8: ...eduction 4 Use the Calibration function to calibrate the handpiece for the AEU 27A Before calibrating ensure that the correct handpiece ratio is selected Press and release the CAL button The following...

Страница 9: ...each Series name These Up Down Arrow buttons can also be used to navigate up or down through the menu options 6 Press the foot pedal to activate motor and begin operation Note that the word RUN displa...

Страница 10: ...rsal S1 SX File 2 ProTaper Universal S2 F1 File 3 ProTaper Universal F2 F3 F4 F5 File 4 ProTaper Universal Retreatment D1 D2 File 5 ProTaper Universal Retreatment D3 Default Series 4 DENTSPLY Tulsa De...

Страница 11: ...CONTROL PANEL FUNCTIONS MOTOR RECEPTACLE LIQUID CRYSTAL DISPLAY CALIBRATION SYSTEM SETUP RECALL FACTORY SETUP AUTO STOP REVERSE REVERSE BEEPER TORQUE WARNING ACCESSORY WARN SLEEP MODE KEYPRESS BEEP S...

Страница 12: ...on on changing these settings B Press and hold the RATIO button to display a list of all the handpiece ratio options Then use the FILE Up Down arrow buttons to point to the desired option and press th...

Страница 13: ...speed accuracy the Ratio Selector must match the ratio of the handpiece being used Note Speed ranges are accurate when the handpieces referenced on page 1 are used with the supplied 30K motor Speed r...

Страница 14: ...allows the motor to run at its maximum torque level When the MAX function is enabled the Auto Stop Reverse ASR function and the torque warning tone are automatically disabled IMPORTANT When a 1 5 han...

Страница 15: ...13...

Страница 16: ...h the system s keypad functions use either 1 Manual Mode or 2 Series Mode below to begin operation 1 MANUAL MODE At any time the user can use the control panel keypad to adjust the ratio speed torque...

Страница 17: ...write the current settings in the selected Series To restore the default Series settings to this Series either re enter the previous settings manually or Recall the Factory Setup see page 16 No Return...

Страница 18: ...nd are not saved into any specific Series when programming B Auto Stop Rev Reverse There are three different Auto Stop modes to choose from After pressing the SELECT button the following menu will dis...

Страница 19: ...enu Cancel Exits out of the menu and returns to the Setup Options menu without saving any setup changes D Torque Warning When the Torque Warning feature is active ON a chirping sound will be audible w...

Страница 20: ...utes of inactivity Sleep Mode 30 will be displayed on the Setup Options menu Cancel Exits out of the Sleep Mode menu and returns to the Setup Options menu without saving any changes G Keypress Beep Wh...

Страница 21: ...e changed then saved altogether at one time This Save Settings option requires the user to either save or ignore the new Setup settings or to cancel out of the menu When this option is selected the fo...

Страница 22: ...t SERIES Arrow buttons to move the flashing cursor laterally under a specific letter in the name then press the FILE Up Down Arrow buttons to select the desired alphanumeric symbol The Down Arrow move...

Страница 23: ...ries Loads the entire file series into the SERIES location Load Individual File Displays a sub menu listing of all the available individual files in the selected SERIES Move the FILE Up Down selector...

Страница 24: ...Returns to the previous Modify Series menu without saving any changes Edit When Select button is pressed the following menu will display example Only A flashing cursor will appear under the first lett...

Страница 25: ...Series menu Move Down When the Select button is pressed the File is immediately moved down one position in the individual file list in the Modify Series menu c End Of Series When the End Of Series opt...

Страница 26: ...wn Arrow buttons to select the desired alphanumeric symbol SELECT To Save Edits After you have edited the Series Name you can Save your new Series by pressing the SELECT button The following Menu will...

Страница 27: ...Turn the Main Power Switch on the back panel On 5 The AEU 27A Display will show the following message Press the RATIO key on the Control Panel 6 The Display will show the following message Press the...

Страница 28: ...per operation and service life of the handpiece Follow the instructions provided with the hand piece for complete maintenance instructions IMPORTANT Protect motor from excess oil draining from handpie...

Страница 29: ...f the console may be cleaned by wiping with a soft cloth moistened with a mild detergent or a 1 10 bleach solution 1 part household bleach to 10 parts water IMPORTANT Use of other cleaning or disinfec...

Страница 30: ...tion Check foot pedal connection Depress foot pedal Increase RPM Increase Torque setting Check that a file is properly seated in the handpiece and the latch is closed Check for dirty under lubricated...

Страница 31: ...nlet is the mains disconnect means Environmental Conditions Operating Temperature 10 to 28 C 50 to 82 4 F Transportation Storage Temperature 20 to 60 C 4 to 140 F Relative Humidity 10 to 90 non conden...

Страница 32: ...s in the Fuse holder Replacement Fuses 1 6A 250V slo blow fuse Fuse size 5 x 20mm Turn the power off and unplug the unit before following the steps below WARNING Fig 10 Fuses NOTE The AEU 27A features...

Страница 33: ...zed European Representative Protection Against Dripping Water Protective Earth Ground 250V Temperature Limitation Atmospheric Pressure Limitation Humidity Limitation Attention consult accompanying doc...

Страница 34: ...32 NOTES...

Страница 35: ...ter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing...

Страница 36: ...P N 421029 Rev C ECO 14684 03 2019 PRINTED IN THE USA P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...

Отзывы: