Flow meter with filter
Rinse the filter of the flow meter regularly.
Testing the pH probe
If the probe meet following qualifications, it can be used in the system and it
is functional:
Probe has no visible mechanical damage.
Measured pH value is in tol/- 1,0 (example - water pH is 7,2 and probe
is measuring 7,9 - the tolerance is 0,7 so lower than 1,0 - the probe is OK)
Probe response to positive or negative changes in water or buffer. (example
- if you dive the probe with dry and clean tip to 7,0 pH buffer the 1 minute
response must be at least 90%)
Testing the REDOX probe
If the probe meet following qualifications, it can be used in the system and it
is functional:
Probe has no visible mechanical damage.
The redox probe naturally ages so its sensitivity but it should never exceed
the limit tolerance -12% At the buffer test 475 mV it should not measure less
than 420 mV.
Probe response to positive or negative changes in the water free chlorine
concentration.
There is no manufacturer of pH and REDOX probes that cover its
products with warranty. ASEKO has decided to cover supplied probes
to its clients by two year warranty period that cover free repair of
supplied probe that will have higher tolerance than above described.
Testing CLF probe
The free chlorine probe should have an output signal of
at least 20 mV
at
a
free chlorine concentration of 0.8 mg / liter
. If the signal is lower, the
probe must be sent for inspection. If the probe has a sufficient signal, it is a
good idea to perform a test with clean water
(the water must stand for 24
hours)
. In pure non-chlorinated water, the signal must be lower than 20 mV.
Otherwise, the probe must be sent for inspection.
pH - Buffer 7,00 #12065
Redox Buffer 475 mV #12063