Troubleshooting
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 36 of 43
Troubleshooting
11.0
Contact Arthrex Technical Support immediately in the event of any damage or
malfunction of the equipment. Attempt the remedies listed in Table 15, in the order in
which they are presented.
Table 15
Common Problems and Remedies
Message
Cause
** Tubing Out **
1.
Tubing sensor indicator not seated.
2.
Ensure that the tubing pressure plug is seated completely.
3.
Change tubing.
4.
If failure remains, return to Arthrex for repair.
Console fails Self
Diagnostic Test
1.
Ensure no tubing is connected to the pump during power on sequence.
2.
If failure remains, return to Arthrex for repair.
Console will not power
up
1.
Check AC mains power cord.
2.
Try alternate power outlet.
3.
Check AC mains’ fuses.
4.
If failure remains, return to Arthrex for repair.
Inadequate distention,
liquid bloody or cloudy
1.
Activate Lavage mode.
2.
Increase pressure.
3.
Decrease outflow.
4.
If failure remains, return to Arthrex for repair.
Does not pump when
activated
1.
Check for error messages.
2.
Open all tubing clamps and shut-off valves.
3.
Ensure actual pressure is below target pressure.
4.
Check if the tubing is pinched, kinked, or blocked.
5.
If failure remains, return to Arthrex for repair.
** Door Open **
1.
Roller Housing not secured – ensure locking lever is properly secured.
2.
If failure remains, return to Arthrex for repair.
Inadequate pressure
1.
Increase pressure.
2.
Activate Lavage mode.
3.
Reduce outflow.
4.
4. Use high-flow cannulas.
5.
If failure remains, return to Arthrex for repair.
No (or inadequate) flow
1.
Check for error messages.
2.
Check that all tubing clamps are open.
3.
Check the settings for flow and pressure.
4.
Check if the tubing is pinched, kinked or blocked.
5.
Check that tubing seats correctly over the rollers.
6.
Verify use of high-flow cannulas.
7.
If failure remains, return to Arthrex for repair.
** Overpressure **
1.
Increase or open outflow.
2.
Manipulate joint to stress-free position.
3.
If failure remains, return to Arthrex for repair.